Siemens Healthcare Diagnostics, Inc: ADVIA Centaur Systems VitD 500 test, Ready Pack, Catalog Number: 10631021. Intended for the invitro diagnostic use in quantitative determination of various assays on the ADVIA Centaur and ADVIA Centaur XP analyzers.

Recall number
Z-1657-2014
Recalling firm
Siemens Healthcare Diagnostics, Inc
Product
ADVIA Centaur Systems VitD 500 test, Ready Pack, Catalog Number: 10631021. Intended for the invitro diagnostic use in quantitative determination of various assays on the ADVIA Centaur and ADVIA Centaur XP analyzers.
Status
Terminated
Initiated
2014-04-10
Posted
2014-05-23
Terminated
2016-05-27
Reason
Samples containing fluorescein may show interference with the ADVIA Centaur Systems TSH3 Ultra, Vitamin D and the ADVIA Centaur BRAHMS Procalcitonin assays.
Root cause
Device Design
Action
Siemens Healthcare Urgent Medical Device Correction (UMDC) was mailed to all affected Siemens Healthcare Diagnostics customers in the United States on April 11, 2014 for customer receipt on April 14, 2014. This notice informed the customers of fluorescein interference with the ADVIA Centaur Systems TSH3-Ultra and Vitamin D Total assays. This notice instructs customers that the TSH3-Ultra and Vitamin D assays can continue to be used for testing any sample that does not contain fluorescein. However, testing blood drawn from a patient, without allowing clearance of fluorescein is not recommended. This notice should be reviewed with the facility's laboratory or medical director. B. An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on April 10, 2014. This notice informed the customer of fluorescein interference with the ADVIA Centaur Systems TSH3-Ultra , Vitamin D Total and ADVIA Centaur BRAHMS Procalcitonin assays. This notice instructs customers that the TSH3-Ultra, Vitamin D and Procalcitonin assays can continue to be used for testing any sample that does not contain fluorescein. However, testing blood drawn from a patient, without allowing clearance of fluorescein is not recommended. This notice should be reviewed with the facilitys laboratory or medical director. (Note: The ADVIA Centaur BRAHMS Procalcitonin assay is not sold in the US.)
Quantity
12,787 kits
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Afghanistan Algeria Armenia Austria Bahrain Belgium Bulgaria Burkina Faso Canary Islands Croatia Cyprus Czech Republic Denmark Estonia Finland France Germany Great Britain Greece Hungary Iceland Iran Iraq Ireland Israel Italy Jordan Kazakhstan Latvia Lebanon Lithuania Luxembourg Macedonia Malta Netherlands Norway Oman Pakistan Poland Portugal Qatar Romania Russian Fed. Saudi Arabia Serbia Slovakia Slovenia Spain Sweden Switzerland Turkey Unit.Arab Emir. Vatican Yemen.
Lot, serial or model codes
Kit lots ending in/Exp Date: 023 07J uly 2014 024 08 September 2014 025 08 December 2014 026 09 February 2015 027 10 April 2015
510(k) numbers
["K110586"]
PMA numbers
not on file
Product code
MRG
Firm FEI
1219913
Firm city
East Walpole
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register