FDA Medical Device Recalls (openFDA)

ArthroCare Corporation: ArthroCare brand Topaz XL ArthroWand with Integrated Cable Wand, Catalog Number: AC4045-01; Product is manufactured and distributed by ArthroCare Corporation, Sunnyvale, CA - Sterile Surgical Device. Product is indicated for ablation and debridement of tendons in orthopedic procedures.

Recall number
Z-1657-2008
Recalling firm
ArthroCare Corporation
Product
ArthroCare brand Topaz XL ArthroWand with Integrated Cable Wand, Catalog Number: AC4045-01; Product is manufactured and distributed by ArthroCare Corporation, Sunnyvale, CA - Sterile Surgical Device. Product is indicated for ablation and debridement of tendons in orthopedic procedures.
Status
Terminated
Initiated
2008-01-31
Posted
2008-09-16
Terminated
2011-01-08
Reason
Devices for which sterility may be compromised as evidenced by punctures in the plastic tray during shipping. The risk to the patient is a potential infection subsequent to the surgical procedure.
Root cause
Packaging process control
Action
Urgent Product Recall letters (Dated January 31, 2008) were issued to end users of the device (medical professionals) and direct distributors. The letter instructs users to return any affected products.
Quantity
2,129
Distribution
Worldwide Distribution --- USA including states of AL, AR, AZ, CA, CO FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, ME, MN, MO, NC, ND, NJ, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, WI, and countries of Sweden, Australia, Germany, Italy, UK, Austria, Saudi Arabia, Belgium, Portugal, UAE, Kuwait, Singapore, Ireland, Spain, Finland, South Korea, Russia and Vietnam.
Lot, serial or model codes
Lot Numbers:   A603250-A,  A609450-A,  A610250-A,  A615750-A,  A616850-A,  A627860-A,  A627860-B,  A604670-A,  A615870-A,  A620170-A,  A623270-A, and A634670-A.
510(k) numbers
["K053567"]
PMA numbers
not on file
Product code
HRX
Firm FEI
3010813691
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register