FDA Medical Device Recalls (openFDA)
Beckman Coulter Inc.: UniCel DxH 800 Coulter Cellular Analysis System, Part Number 629029. Beckman Coulter, Inc. Brea, CA. Quantitative, automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories.
- Recall number
- Z-1656-2010
- Recalling firm
- Beckman Coulter Inc.
- Product
- UniCel DxH 800 Coulter Cellular Analysis System, Part Number 629029. Beckman Coulter, Inc. Brea, CA. Quantitative, automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories.
- Status
- Terminated
- Initiated
- 2010-03-02
- Posted
- 2010-05-26
- Terminated
- 2012-05-23
- Reason
- Issue 1: The Diagnostic procedure, Simulate-CDNR Cycle, is not functioning properly. Impact: There is a potential for inadequate sample preparation of samples following the use of this Diagnostic function impacting WBC, Differential, Reticulocyte and/or NRBC results. Issue 2: Use of the Auto Prune function causes the deletion of Quality Control files. Impact: If the last run for a control lot
- Root cause
- Software design
- Action
- Beckman Coulter, Inc. issued an "Urgent: Product Corrective Action" notification dated March 10, 2010. Consignees were informed of the affected product and instructed on proper steps to take and requested to return the enclosed acknowledgement form. For further information, contact your local Beckman Coulter Representative or call 1-800-526-7694.
- Quantity
- 63 units
- Distribution
- Nationwide Distribution (AZ, CA, CO, FL, GA, IA, IL, IN, KS, LA, MO, NC, NJ, NV, NY, OH, PA, SC, TN, TX, UT, WA, and WI) and Canada.
- Lot, serial or model codes
- All Serial Numbers
- 510(k) numbers
- ["K081930"]
- PMA numbers
- not on file
- Product code
- GKZ
- Firm FEI
- 2050012
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as