Skip to the content

FDA Medical Device Recalls (openFDA)

Therakos Inc: Therakos CellEx Photopheresis System; Photopheresis Procedural Kits, for use with the THERAKOS CellEx Instrument;

Recall number
Z-1654-2010
Recalling firm
Therakos Inc
Product
Therakos CellEx Photopheresis System; Photopheresis Procedural Kits, for use with the THERAKOS CellEx Instrument;
Status
Terminated
Initiated
2010-03-29
Posted
2010-05-26
Terminated
2013-07-18
Reason
Higher than normal blood leaks on THERAKOS CELLEX Procedural Kits that are intended for use with the THERAKOS CELLEX Photopheresis System
Root cause
Nonconforming Material/Component
Action
On March 29, 2010, Therakos Photopheresis sent Important Product Correction Notification letters by FedEx to all direct end use and distributor customers who received lot X120 in addition to phone contact. On March 31, 2010 those who received lot X123 were sent letters by Fed Ex as well as phone contact.. The one Canadian customer was notified by phone on March 31, 2010 followed by a Fed Ex letter. On April 12, consignees who were shipped Lot X119 were contacted by phone, followed by a hard copy letter. On May 7, 2010 all direct end use and distributor consignees were notified of the additional lot being added to the recall. Questions may be directed to Therakos Customer Technical Services at 1-877-865-6850.
Quantity
280 units in US (1 sales unit contains 3 kits each)
Distribution
Nationwide Distribution, and Canada and Germany
Lot, serial or model codes
CLXUSA (US): Lot X120, Exp 12/10; X123, Exp 1/2011; Lot Y106, Exp 2/1/11. CLSECP (For European Union), Lot X119, Exp 12/2014; Lot X123; Exp 1/2015
510(k) numbers
not on file
PMA numbers
not on file
Product code
LNR
Firm FEI
2523595
Firm city
Raritan
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register