Siemens Medical Solutions USA, Inc: ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.
- Recall number
- Z-1653-2026
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.
- Status
- Open, Classified
- Initiated
- 2026-03-10
- Posted
- 2026-04-07
- Terminated
- not on file
- Reason
- During patient examinations, the system may sporadically display a major increase of the x-ray dose applied to the patient. Despite the reported dose showing significantly higher, the actual applied x-ray dose levels are in accordance with the system settings and imaging conditions, as well as the applicable regulatory requirements.
- Root cause
- Radiation Control for Health and Safety Act
- Action
- Siemens Medical Solutions USA, Inc. has initiated a Customer Advisory Notice, dated March 9, 2026, to all affected customers.
- Quantity
- U.S. 758, OUS 2077
- Distribution
- Worldwide - US Nationwide distribution.
- Lot, serial or model codes
- ARTIS pheno UDI: 4056869046877. ARTIS icono biplane UDI: 4056869063317. ARTIS icono ceiling UDI: 4056869295923. ARTIS icono floor UDI: 4056869149325.
- 510(k) numbers
- ["K241572"]
- PMA numbers
- not on file
- Product code
- OWB
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28