FDA Medical Device Recalls (openFDA)
Roche Diagnostics Operations, Inc.: Factor V Leiden kit catalog number 03610179001, Roche Diagnostics, Indianapolis, IN The Factor V Leiden Kit allows the detection and genotyping of a single point mutation (G to A at position 1691) of the human Factor V gene, referred to as Factor V Leiden mutation, from DNA isolated from human whole peripheral blood. The test is performed on the LightCycler 2.0 Instrument utilizing polymerase chain reaction (PCR) for the amplification of Factor V DNA recovered from clinical specimens and fluorogenic target-specific hybridization for the detection of genotyping of the amplified Factor V DNA. The Factor V Leiden test is an in vitro diagnostic test for the detection and genotyping of the Factor V Leiden mutation as an aid to diagnosis in the evaluation of patients with suspected thrombophilia. The test is intended to be used on the LightCycler 2.0 Instrument using the LightCycler Software 4.05 or 4.1.
- Recall number
- Z-1653-2011
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Product
- Factor V Leiden kit catalog number 03610179001, Roche Diagnostics, Indianapolis, IN The Factor V Leiden Kit allows the detection and genotyping of a single point mutation (G to A at position 1691) of the human Factor V gene, referred to as Factor V Leiden mutation, from DNA isolated from human whole peripheral blood. The test is performed on the LightCycler 2.0 Instrument utilizing polymerase chain reaction (PCR) for the amplification of Factor V DNA recovered from clinical specimens and fluorogenic target-specific hybridization for the detection of genotyping of the amplified Factor V DNA. The Factor V Leiden test is an in vitro diagnostic test for the detection and genotyping of the Factor V Leiden mutation as an aid to diagnosis in the evaluation of patients with suspected thrombophilia. The test is intended to be used on the LightCycler 2.0 Instrument using the LightCycler Software 4.05 or 4.1.
- Status
- Terminated
- Initiated
- 2010-10-26
- Posted
- 2011-03-14
- Terminated
- 2011-08-30
- Reason
- If the bar code is printed from the Batch Results Screen, a wrong batch ID may be printed, while the correct batch ID is still assigned to the batch in the data repository of the MPLC 2.0 software. This may result in sample mis identification. If printed from the ordering screens, the bar code is printed with the correct batch ID.
- Root cause
- Software design
- Action
- The firm, Roche, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated November 04, 2010 to its customers. The letter described the product, problem and action to be taken. The customers were instructed to print batch ID barcode labels from the ordering or stage set up screens only; print Cooling Block barcode labels from the Post Elution Edit screen only; complete and return the attached UMDC Faxback Form for MagNA Pure LC 2.0 Instrument [UMDC 10-228] via fax to 1-877-551-7152 and file this correction notice for future reference. The firm plans to issue a software update in first quarter 2011. If you have questions about the information contained in this UMDC, for Clinical customers, please contact Roche Molecular Diagnostics Technical Service at 1-800-526-1247; for Research customers, please contact Roche Applied Sciences Technical Service at 1-800-262-4911.
- Quantity
- not on file
- Distribution
- Worldwide distribution: AL, AZ, CA, CO, DC, FL, GA, HI, IL, KY, LA, MA, MD, ME, MN, MO, MT, NC, ND NE, NH, NJ, NM, NY, OH, OK, OR, SC, SD, TN, TX, UT, VA, VT, WI and Puerto Rico; and country of Europe.
- Lot, serial or model codes
- n/a
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- NPQ
- Firm FEI
- 1823260
- Firm city
- Indianapolis
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28