Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic MiniMed, Inc.: InPen smart insulin pen, REF: MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA, MMT-105ELBLNA, MMT-105ELGYNA, MMT105ELPKNA

Recall number
Z-1650-2025
Recalling firm
Medtronic MiniMed, Inc.
Product
InPen smart insulin pen, REF: MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA, MMT-105ELBLNA, MMT-105ELGYNA, MMT105ELPKNA
Status
Open, Classified
Initiated
2025-03-04
Posted
2025-04-21
Terminated
not on file
Reason
Insulin pens may have been incorrectly assembled therefore users could experience the following: either the insulin cartridge will not fit into the cartridge holder; or the cartridge holder may be difficult to remove from the insulin pen. If the issue is present, and if the user does not have backup insulin they may experience temporary hyperglycemia.
Root cause
Mixed-up of materials/components
Action
On 3/4/2025, correction notices were sent to customers and distributors informing them of the following: 1) If you have already inserted your insulin cartridge into your device without difficulty removing the cartridge holder, then you can continue normal use. 2) If you have a device from one of the affected lot numbers and have not yet started using it, or have had difficulties in removing the cartridge holder or inserting the insulin cartridge, please call 1-800-646-4633, option 1, or go to https://info.medtronicdiabetes.com/ICH-replacement-form to request a replacement product. Do not use this device and dispose of it according to local regulations. 3) Complete and return the confirmation form via link: https://info.medtronicdiabetes.com/cartridge-holder Distributors are asked to notify impacted customers. If you have further questions or need assistance, please call the firm's 24-Hour Technical Support line at 1-800-646-4633, option 1.
Quantity
8,986
Distribution
US: OK, IA, NC, SC, MI, MN, CA, AZ, AL, MO, WA, IL, KS, TX, FL, MT, MD, OH, VA, NY, TN, HI, MA, NJ, PA, CO, WI, GA, CT, NE, OR, LA, NM, SD, ND, IN, NV, WV, RI, AR, NH
Lot, serial or model codes
REF/UDI-DI(GTIN)/Lot: MMT-105NNBLNA/763000827182/D0020, D0034, D0036, D0038, D0045, D0046, D0050, D0057; MMT-105NNGYNA/763000827199/D0026, D0037, D0039, D0041, D0042, D0043, D0049; MMT-105NNPKNA/763000827205/D0029, D0030, D0040, D0051, D0062; MMT-105ELBLNA/763000827151/D0021, D0033, D0035, D0044, D0059, D0060, D0061; MMT-105ELGYNA/763000827168/D0031, D0032, D0047, D0068; MMT105ELPKNA/763000827175/D0027, D0028, D0048
510(k) numbers
["K160629","K190487","K192841","K201337","K242775"]
PMA numbers
not on file
Product code
FMF
Firm FEI
3003166194
Firm city
Northridge
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register