FDA Medical Device Recalls (openFDA)
Smiths Medical ASD, Inc.: SW-2008 Snuggle Warm Sterile Cardiac Connective Warming Blanket; used with the Equator Convective Warmer, are intended for thermal regulation of a patient's temperture and are designed to prevent hypothermia and/ or reduce cold discomfort during and after surgical procedures.
- Recall number
- Z-1650-2014
- Recalling firm
- Smiths Medical ASD, Inc.
- Product
- SW-2008 Snuggle Warm Sterile Cardiac Connective Warming Blanket; used with the Equator Convective Warmer, are intended for thermal regulation of a patient's temperture and are designed to prevent hypothermia and/ or reduce cold discomfort during and after surgical procedures.
- Status
- Terminated
- Initiated
- 2014-04-22
- Posted
- 2014-05-22
- Terminated
- 2015-02-27
- Reason
- The hose could separate from the blanket.
- Root cause
- Process change control
- Action
- On April 24, 2014, consignees were notified of the recall via letter. They were instructed to do the following: Inspect inventory and segregate any unused Blankets listed in the notification, complete and return the attached Confirmation Form by fax or e-mail within 7 days of receipt of this notice, and destroy the product.
- Quantity
- 2120 devices
- Distribution
- Worldwide Distribution Including the States of CA, IL, KY, MS, NC, TN, TX, WA, WI and the Countries of BH, DE, NO, RU
- Lot, serial or model codes
- Lot: 1006733, 2474599, 2488138, 2523531, 2536250, 2558180, 2563033, 2564447, 2564448, 2615712
- 510(k) numbers
- ["K040632"]
- PMA numbers
- not on file
- Product code
- DWJ
- Firm FEI
- 1221261
- Firm city
- Rockland
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28