FDA Medical Device Recalls (openFDA)
Philips Medical Systems (Cleveland) Inc: Brilliance CT 64. Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
- Recall number
- Z-1649-2015
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Product
- Brilliance CT 64. Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
- Status
- Terminated
- Initiated
- 2015-03-16
- Posted
- 2015-05-22
- Terminated
- 2017-02-22
- Reason
- The firm was informed that while raising the patient couch on the system to perform an exam, the couch unexpectedly descended to the lowest point without being commanded to do so.
- Root cause
- Employee error
- Action
- On 3-18-2015 the firm sent Urgent Field Safety Notices to their customers.
- Quantity
- 153 units total
- Distribution
- Distributed in China.
- Lot, serial or model codes
- System Code #, 728231, Serial #, 9568, 9758, 10097, 10143, 10267, 10269, 10275, 10427, 10436, 10521, 10545, 10579, 10622, 90097, 90218, 95203, 95319, 95328, 95362, 95369, 95391, 95414, 95427, 95433, 95471, 95486, 95544, 95586, 95604, 95605, 95617, 95628, 95632, 95640, 95672, 95752, 95754 & 95757.
- 510(k) numbers
- ["K033326"]
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 1525965
- Firm city
- Cleveland
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28