Siemens Medical Solutions USA, Inc: Artis pheno- Model No. 10849000

Recall number
Z-1646-2022
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Artis pheno- Model No. 10849000
Status
Open, Classified
Initiated
2022-08-10
Posted
2022-08-26
Terminated
not on file
Reason
Siemens has become aware of a potential hardware issue with Artis Q, Artis Q.zen, Artis pheno and Artis icono systems with a specific lot of X-ray tubes. In rare cases, if the error detection mechanism fails, it will not be possible to release X-ray any longer until system shutdown.
Root cause
Device Design
Action
On August 10, 2022, the firm notified customers via "URGENT: MEDICAL DEVICE CORRECTION" Letter. Customers were advised that if the error occurs, the user should try to recover normal operation by a shutdown and restart of the system. Siemens will correct the error with a hardware modification via Update Instruction AX028/22/S and AX039/22/S. Siemens' service organization will contact customers shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. If you have any further questions, call 610-219-4834 or email: meredith.adams@siemens-healthineers.com.
Quantity
1,389 (590 US; 799 OUS) (Total)
Distribution
Worldwide Distribution: US (nationwide) and OUS (Foreign) to countries of: United Arab Emirates; Armenia; Angola; Argentina; Austria; Australia; Bangladesh; Belgium; Bulgaria; Bahrain; Brazil; Canada; Switzerland; China; Colombia; Costa Rica; Czech Republic; Germany; Denmark; Dominican Republic; Algeria; Estonia; Egypt; Spain; Finland; France; United Kingdom; Greece; Guatemala; Hong Kong; Croatia; Inhouse Company; Indonesia; Ireland; Israel; India; Iran; Italy; Jordan; Japan; South Korea; Kuwait; Kazakhstan; Lebanon; Luxembourg; Monaco; Mongolia; Martinique; Mexico; Malaysia; Netherlands; Norway; New Zealand; Oman; Philippines; Poland; Portugal; Qatar; Serbia; Russian Federation; Saudi Arabia; Sweden; Singapore; Slovenia; Slovakia; El Salvador; Thailand; Turkey; Taiwan; Ukraine; Vietnam and South Africa.
Lot, serial or model codes
UDI-DI: 04056869046877 Serial Numbers 164766 164140 164138 164287 164740 164210 164183 164184 164776 164312 164071 164034 164391 164763 164306 164735 164277 164095 164072 164771 164743 164048 164741 164241 164186 164069 164213 164728 164263 164060 164113 164398 164366 164279 164774 164074 164084 164086 164098 164153 164330 164389 164394 164772 164718 164163 164297 164778 164333 164377 164397 164300 164105 164227 164022 164250 164739 164703 164187 164211 164100 164097 164775 164303 164352 164726 164724 164760 164371 ***Updated 9/23/22*** 164794 164400 ***Updated 1/9/23*** 164793 164172 164795
510(k) numbers
["K220409"]
PMA numbers
not on file
Product code
OWB
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register