Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Healthcare Informatics, Inc.: Philips XRE, the dictation editor system included in the Philips XIRIS 8.1.50 XIRIS is a radiology information system (RIS) is a computer system that assist radiology services in the storing, manipulating and retrieving of patient information. This included patient scheduling and tracking and workflow and reporting management.

Recall number
Z-1646-2014
Recalling firm
Philips Healthcare Informatics, Inc.
Product
Philips XRE, the dictation editor system included in the Philips XIRIS 8.1.50 XIRIS is a radiology information system (RIS) is a computer system that assist radiology services in the storing, manipulating and retrieving of patient information. This included patient scheduling and tracking and workflow and reporting management.
Status
Terminated
Initiated
2014-05-05
Posted
2014-05-21
Terminated
2015-04-28
Reason
When using XIRIS Rio version (8.1.50.0) in conjuncture with iSite PACS a synchronization error can occur and if not noticed, information for the incorrect patient may be displayed
Root cause
Software design
Action
Philips Healthcare sent an Urgent Field Safety Notice dated May 5, 2014. to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were advised of the synchronization problem and request that a copy of this notification be kept with the equipment instructions for use. The document goes on to outline the affected product, the problems and conditions when it can occur. A photo of the label for easy identity the actions to be taken are provided. Philips is informing users of the issue via the notification and will deploy a software correction to all customers when available. Questions may be directed to Philips reps or Care at 1-866-946-4701 or OneEMSProd@philips.com
Quantity
5 units
Distribution
Worldwide Distribution - US including the states of FL and TX and the countries Canada and South Africa.
Lot, serial or model codes
Model XIRIS 8.1 version XIRIS 8.1.50.0
510(k) numbers
not on file
PMA numbers
not on file
Product code
OUG
Firm FEI
3002971142
Firm city
Foster City
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register