Philips North America Llc: EarlyVue VS30 Vitals monitor with Software Versions A.00.02 and A.00.01 Product Number: 863380

Recall number
Z-1645-2023
Recalling firm
Philips North America Llc
Product
EarlyVue VS30 Vitals monitor with Software Versions A.00.02 and A.00.01 Product Number: 863380
Status
Open, Classified
Initiated
2023-03-06
Posted
2023-05-26
Terminated
not on file
Reason
EarlyVue VS30 Vital Signs Monitor and EarlyVue VS30 Vitals monitor Missing Calibration Alarm, a risk of an inaccurate CO2 measurement which may lead to a failure to recognize a change in patient condition.
Root cause
Software design
Action
Philips issued Urgent Field Safety Notice Letter on 3/3/23. Letter states reason for recall, health risk and action to take: 1. Identify the last calibration date of the device and review your hospital maintenance plan. Determine if your device has E01 End Tidal CO2 and has not been calibrated within one year of the production calibration. If needed adjust the maintenance plan and incorporate the calibration of your affected device. a. The device should not be used until the calibration or maintenance plan has been reviewed. b. The calibration date can be accessed via the following steps on the device: Settings > Admin > Diagnostics > Page 2 of Diagnostics > Maintenance > 129 > CO2 Test. The image below shows the example of the screen to see the last calibration date of your device: 2. Review this URGENT Field Safety Notice in its entirety and pass this notice to all those who need to be aware within your organization or to any organizations where the potentially affected devices have been transferred. (If appropriate). 3. Complete and return attached form to Philips promptly to confirm receipt of the URGENT Field Safety Notice Letter, understanding of the issue, and required actions to be taken. 4. Actions planned by Philips Hospital Patient Monitoring to correct the problem A Philips representative will reach out to you to arrange a software upgrade to your monitor(s). Philips will start to schedule the upgrade upon formal release of the field action, which is planned for a six month implementation timeline. If you need any further information or support concerning this issue, please contact your local Philips representative
Quantity
1421 units US; 3 units OUS
Distribution
Nationwide Foreign: Belgium Brazil Canada Denmark Germany Guam Hong Kong Indonesia Italy Norway Poland Singapore Sweden Switzerland United Kingdom
Lot, serial or model codes
UDI: (01)00884838091412 Software Version A.00.02 and A.00.01
510(k) numbers
["K190624"]
PMA numbers
not on file
Product code
DSJ
Firm FEI
3016618143
Firm city
Cambridge
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register