Beckman Coulter Inc.: Access Hybritech Free PSA, Part Number: 37210 Free PSA is used to aid in the differentiation between prostate cancer and benign prostatic conditions when total PSA levels are between 4 and 10 ng/mL.
- Recall number
- Z-1645-2012
- Recalling firm
- Beckman Coulter Inc.
- Product
- Access Hybritech Free PSA, Part Number: 37210 Free PSA is used to aid in the differentiation between prostate cancer and benign prostatic conditions when total PSA levels are between 4 and 10 ng/mL.
- Status
- Terminated
- Initiated
- 2011-10-11
- Posted
- 2012-05-22
- Terminated
- 2013-03-18
- Reason
- A recall was initiated because Beckman Coulter has confirmed a change to the recommended diluents for the Access Hybritech free PSA reagent.
- Root cause
- Process design
- Action
- Beckman Coulter sent a Product Correction letter dated October 11, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed regarding a change to the recommended diluents for the Access Hybritech free PSA reagent. Patient safety is not affected by this change. " Do not dilute elevated free PSA patient samples with Wash Buffer II. " Do not dilute elevated free PSA patient samples using the UniCel DxI onboard dilution feature. " Dilute elevated free PSA samples with Access Hybritech free PSA Calibrator S0 only. " A review of historical patient results is not required because there is no effect on clinical interpretation. * The UniCel DxI onboard dilution feature is only available in limited geographies. " An updated version of the UniCel DxI Assay Protocol File (APF) will remove the Access Hybritech free PSA (dfPSA) onboard dilution assay from the list of available assays. " The Access Hybritech free PSA IFU will be updated to remove Wash Buffer II as an acceptable diluent for elevated patient samples. Correct any procedure manuals or other documents created in your laboratory that list Wash Buffer II as an acceptable diluent for the Access Hybritech free PSA assay. Please share this information with your laboratory staff. If you have forwarded any of the affected product listed above to another laboratory, please provide a copy of the letter to them. Please complete and return the enclosed response form within 10 days so we are certain you received this important communication. If you have any questions regarding this notification, please contact our Customer Support Center " Via our website, http://www.beckmancoulter.com/customersupport/support " Via phone, call 1-800-854-3633 in the United States and Canada Outside the United States and Canada, contact your local Beckman Coulter Representative. We apologize for the inconvenie
- Quantity
- 90,965 units total (24,951 units in US)
- Distribution
- Worldwide Distribution -- USA (nationwide ) and the countries of Algeria, Angola, Argentina, Armenia, Australia, Austria, Bangladesh, Belgium, Bosnia, Herzegovina, Brazil, Bulgaria, Canada, China, Colombia, Costa Rica, Croatia, Czech Republic, Djibouti, Ecuador, Egypt, El Salvador, Ethiopia, Finland, France, French, Polynesia, Georgia, Germany, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Lebanon, Libyan, Arab, Jamahiriya, Macao, Mexico, Moldova, Republic of Morocco, Myanmar, Namibia, Netherlands, New Zealand, Nigeria, Oman, Panama, Philippines, Poland, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab, Emirates, United Kingdom, Uruguay ,Venezuela, Viet Nam, Yemen and Zimbabwe.
- Lot, serial or model codes
- All lots
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- MTG
- Firm FEI
- 2050012
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28