Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Communications Corp: Ceiling Suspensions (Surgical Light and Flat Panel Suspensions); Catalog Numbers: 0682000863, 06820001009, 0682001010, 0682001076, 0682000864, 0682000857, 0682000858, 0682001077, 0682000865, 0682001011, 0682001078, 0682000866, 0682001012, 0682001080, 0682000867, 0682000959, 0682001079, 0682000850, 0682000851, 0682000852, 0682001269, 0682001567, 0682001569, 0682001570, 06820012691, 06820015671, 06820015691, & 0680015701. The ceiling suspensions are available in different configurations; they can hold monitors (flat panel suspensions), cameras (navigation suspensions arms) or surgical lights. The flat panel could have fixed or adjustable monitor yokes and may be single or dual arm configuration. Each ceiling suspension could hold a maximum of two surgical lights and either one camera or one monitor. There may be multiple ceiling/boom suspensions in an operating room. Intended for use as ceiling-mounted devices for supporting or positioning surgical lights, cameras, and/or monitors.

Recall number
Z-1645-2010
Recalling firm
Stryker Communications Corp
Product
Ceiling Suspensions (Surgical Light and Flat Panel Suspensions); Catalog Numbers: 0682000863, 06820001009, 0682001010, 0682001076, 0682000864, 0682000857, 0682000858, 0682001077, 0682000865, 0682001011, 0682001078, 0682000866, 0682001012, 0682001080, 0682000867, 0682000959, 0682001079, 0682000850, 0682000851, 0682000852, 0682001269, 0682001567, 0682001569, 0682001570, 06820012691, 06820015671, 06820015691, & 0680015701. The ceiling suspensions are available in different configurations; they can hold monitors (flat panel suspensions), cameras (navigation suspensions arms) or surgical lights. The flat panel could have fixed or adjustable monitor yokes and may be single or dual arm configuration. Each ceiling suspension could hold a maximum of two surgical lights and either one camera or one monitor. There may be multiple ceiling/boom suspensions in an operating room. Intended for use as ceiling-mounted devices for supporting or positioning surgical lights, cameras, and/or monitors.
Status
Terminated
Initiated
2010-04-15
Posted
2010-05-21
Terminated
2011-01-24
Reason
Installation records are incomplete.
Root cause
Employee error
Action
An Urgent: Medical Device Product Correction customer notification letter, dated 4/15/2010, has been sent out to all affected accounts via regular mail in the form of a certified, return request receipt letter. Accounts will receive a follow-up phone call from a Stryker Representative to schedule a re-inspection of equipment to see if it mets established critical specifications or not. The firm recommends that customers continue using all Stryker Communications products listed in the notice. Questions can be directed to Katie Blackwell at 972-410-7660 or via email at StrykerCommunicationsRA@stryker.com.
Quantity
not on file
Distribution
Nationwide Distribution
Lot, serial or model codes
Product installed Between January 1, 2008 and August 31, 2009.
510(k) numbers
["K031068","K033132"]
PMA numbers
not on file
Product code
BRY
Firm FEI
3002975342
Firm city
Flower Mound
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register