Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott: Amplatzer TorqVue LP Catheter (TVLPC), Model no. 9-TVLPC4F90/080

Recall number
Z-1644-2025
Recalling firm
Abbott
Product
Amplatzer TorqVue LP Catheter (TVLPC), Model no. 9-TVLPC4F90/080
Status
Open, Classified
Initiated
2025-03-17
Posted
2025-04-21
Terminated
not on file
Reason
Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may lead to prolonged procedure, blood loss, or air ingress with potential for air embolism.
Root cause
Process control
Action
On March 17, 2025, the firm notified customers via hand-delivered customer communications titled "Urgent Medical Device Recall." Customers were instructed to share the notice with applicable personnel and to return any remaining unused devices from impacted lots. Abbott representatives will assist with product return and replenishment. If you have any questions, please contact your local Sales Representative or Abbott Support at 1-800-544-1664.
Quantity
2683
Distribution
Worldwide - US Nationwide distribution.
Lot, serial or model codes
UDI-DI 00811806012926 Lots 9078966 9093895 9134623 9144113 9148110 9148579 9155868 9158302 9241591 9241593 10004059 10011461 10011462 10011463 10011465 10017790 10017792 10196106 10196119 10198795 10198801 10198889 10207472 10207503 10274454 10283818
510(k) numbers
["K162228"]
PMA numbers
not on file
Product code
DQY
Firm FEI
3005334138
Firm city
Plymouth
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register