Skip to the content

FDA Medical Device Recalls (openFDA)

Mako Surgical Corporation: RESTORIS MCK TIBIAL BASEPLATE-RM/LL used in unicompartmental knee arthroplasty and knee replacements.

Recall number
Z-1644-2018
Recalling firm
Mako Surgical Corporation
Product
RESTORIS MCK TIBIAL BASEPLATE-RM/LL used in unicompartmental knee arthroplasty and knee replacements.
Status
Terminated
Initiated
2017-12-01
Posted
not on file
Terminated
2020-12-14
Reason
Incorrect product and/or label. Sizing is labeled incorrectly.
Root cause
Error in labeling
Action
An URGENT MEDICAL DEVICE RECALL was sent on December 05, 2017 Product Field Action #: 1658081 Description: MCK TIBIAL BASEPLATE-RM/LL-SZ 2 and MCK TIBIAL BASEPLATE-RM/LL-SZ 7 Item No.: 180612; Lot Number: 26080317-01180617; and Lot Number: 26150217-01 Dear Customer, Stryker has initiated an urgent, voluntary, lot-specific recall for the MCK TIBIAL BASEPLATE-RM/LL-SZ 2 and MCK TIBIAL BASEPLA TE-RM/LL-SZ 7 referenced above. The intent of this letter is to inform you of the product recall that was initiated on December 01, 2017 by Stryker, and to list known hazards potentially associated with the use of the above referenced products and list the risk mitigation factors. Stryker has discovered that the packaging of certain sizes and lots of the above-referenced product may contain the incorrect product and/or label. Two reports were received with the product/label discrepancy. In one report, the labeling of the implant box outer label stated Size 2 RM/LL, and the labeling of the implant sticker (Patient label) located inside the outer box stated Size 7 RM/LL. The correct implant Size 2RM/LL was inside the box. The patient label was incorrect in this report. The second report described that a size 2 implant was in a box labeled as a size 7 implant. Potential Hazards Technical and medical assessments are currently underway to determine any potential hazards associated with the use of the product. Additional communication will be forwarded upon completion of the internal investigation on this issue. Risk Mitigation The difference in Size 2 RM/LL implant and Size 7 RM/LL implant is easily identified by the end user and would not likely be implanted as the discrepancy would be obvious. Actions Needed 1. Please inform users of this Urgent Product Recall and forward this notice to all individuals who need to be aware within your organization 2. Hospitals/Branches/Agencies: Complete and sign the enclosed Recall Notification Business Reply Form and
Quantity
126
Distribution
AL, AR, AZ, CA, CO, FL, GA, ID, IL, IN, KS, LA, MA, MD, MI, MN, NC, NE, NH, NJ, NY, OK, OR, PA, RI, SC, TX, UT, VA, WA, WY, and Hawaii Australia, Italy, Netherlands, Singapore, South Korea, United Kingdom, and Vietnam
Lot, serial or model codes
Size 2 Item # 180612 Lot # 26080317-01 and  Size 7 Item # 180617 Lot # 26150217-01.
510(k) numbers
["K170584","K170891"]
PMA numbers
not on file
Product code
OLO
Firm FEI
3005985723
Firm city
Davie
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register