Computerized Medical Systems Inc: XiO Radiation Treatment Planning System, Versions 4.30.00-4.64.00. Used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed. The system will calculate and display, both on-screen and in hard-copy, either two- or three-dimensional radiation dose distributions within a patient for a given treatment plan set-up.

Recall number
Z-1644-2012
Recalling firm
Computerized Medical Systems Inc
Product
XiO Radiation Treatment Planning System, Versions 4.30.00-4.64.00. Used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed. The system will calculate and display, both on-screen and in hard-copy, either two- or three-dimensional radiation dose distributions within a patient for a given treatment plan set-up.
Status
Terminated
Initiated
2012-04-30
Posted
2012-05-22
Terminated
2014-04-01
Reason
XiO will underestimate the dose delivered when the closed leaf gap is set to 0mm on XiO and a value other than 0mm is set on the accelerator.
Root cause
Software design
Action
Computerized Medical Systems, Inc. sent an Important Safety Notice the week of April 30, 2012, by regional support managers via e-mail, fax, or regular mail to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. North American customers were instructed to go to www.elekta.com, select the SupportPlus Login at the top of the page, enter their portal, select Downloads/Updates to download the latest patches. For Customers outside North America, go to www.elekta.com, select the SOFTWARE button on the left, Treatment Planning Software, then Contact TPS Support tab, Download TPS Software Patches. ............................................................. Customers were also instructed to distribute this notice to any and all users of CMS software at their organization who are potentially affected by this issue. For questions regarding this recall call 408-380-8023.
Quantity
1,652 sites received the versions
Distribution
Worldwide Distribution - USA (nationwide) and the country of Canada
Lot, serial or model codes
Versions 4.30.00, 4.31.00, 4.32.00, 4.33.00, 4.33.01, 4.33.02, 4.34.00, 4.34.01, 4.34.02, 4.40.00, 4.40.04, 4.40.05, 4.41.00, 4.41.01, 4.41.02, 4.42.00, 4.43.00, 4.44.01, 4.44.02, 4.44.03, 4.44.04, 4.45.00, 4.46.00, 4.46.01, 4.47.00, 4.50.00, 4.50.01, 4.50.02, 4.50.03, 4.50.04, 4.51.00, 4.51.01, 4.51.02, 4.60.00, 4.60.01, 4.61.00, 4.62.00, 4.62.01, 4.62.02, 4.62.03, 4.62.04, 4.62.05, 4.62.06, 4.63.00, and 4.64.00.
510(k) numbers
K102216
PMA numbers
not on file
Product code
MUJ
Firm FEI
1937649
Firm city
Maryland Heights
Firm state
MO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source