Skip to the content

FDA Medical Device Recalls (openFDA)

Olympus Corporation of the Americas: Brand Name: Olympus PKS Cutting Forceps w/Cord, 5mm 24cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920000PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the PKS electrical generator.

Recall number
Z-1643-2026
Recalling firm
Olympus Corporation of the Americas
Product
Brand Name: Olympus PKS Cutting Forceps w/Cord, 5mm 24cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920000PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the PKS electrical generator.
Status
Open, Classified
Initiated
2026-02-25
Posted
2026-03-25
Terminated
not on file
Reason
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Root cause
Nonconforming Material/Component
Action
On February 25, 2026 URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: 1. Examine inventory and quarantine any affected devices 2. Cease usage of the product with immediate effect 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization 4. Even if you no longer have the device, Olympus requests that you acknowledge receipt of this letter through the Olymps portal 5. Please forward this notice to other users who may have the affected products if you have further distributed it.
Quantity
7 units
Distribution
Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
Lot, serial or model codes
Model: 920000PK; UDI: 00821925038080; Lot Numbers: All unexpired;
510(k) numbers
["K023492"]
PMA numbers
not on file
Product code
GEI
Firm FEI
2429304
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register