FDA Medical Device Recalls (openFDA)
DeRoyal Industries Inc: DeRoyal(R) Medium Specimen Retrieval Bag, REF 5500-000-003, Rx Only Product Usage: The device is a single-use specimen retrieval bag indicated for use in laparoscopic procedures to capture and remove organs or tissue from the body cavity.
- Recall number
- Z-1642-2015
- Recalling firm
- DeRoyal Industries Inc
- Product
- DeRoyal(R) Medium Specimen Retrieval Bag, REF 5500-000-003, Rx Only Product Usage: The device is a single-use specimen retrieval bag indicated for use in laparoscopic procedures to capture and remove organs or tissue from the body cavity.
- Status
- Terminated
- Initiated
- 2015-04-17
- Posted
- 2015-05-21
- Terminated
- 2015-12-29
- Reason
- The firm received reports of specimen retrieval bags tearing, ripping, and coming apart during use.
- Root cause
- Device Design
- Action
- DeRoyal sent an Urgent: Voluntary Recall letter dated April 17, 2015 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to 1. Check your inventory for the product and specific lot number from the attached Recall Response Form. 2. If you have affected inventory, please destroy it and dispose of it according to your facilitys procedures. 3. Record all information on the included Recall Response and Notice of Destruction forms. 4. If you purchase through a distributor, please indicate their name/location on the Notice of Destruction Form so that we may issue credit appropriately. 5. Return both forms to recalls@deroyal.com or 865-362-3716 (fax) even if you no longer have affected inventory. 6. Once we have received your Recall Response and Notice of Destruction forms, you will receive credit for destroyed product as indicated on the form. 7. If you are a distributor of these products, please forward this recall notice on to your end-users, or provide us with your customer listing and we will gladly contact them for you. For questions regarding this recall, please contact your DeRoyal Sales Representative or DeRoyal Customer Service at 1-800-251-9864.
- Quantity
- 5359 units
- Distribution
- US Nationwide Distribution
- Lot, serial or model codes
- 1211QDA403, 1301QDA402, 1301QDA410, 1303QDA401, 211QDA403, 32046564, 32119911, 32338381, 32562251, 32783812, 32886441, 32891051, 32899882, 32942609, 33114274, 34939294, 35229930, 35579999, 35738515
- 510(k) numbers
- ["K132375"]
- PMA numbers
- not on file
- Product code
- GCJ
- Firm FEI
- 1043214
- Firm city
- Powell
- Firm state
- TN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28