Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Howmedica Osteonics Corp.: Restoration Modular Cone Bodies; titanium Alloy and Hydroxyapatite, Sterile. Catalog number 6276-1-021, Catalog number 6276-1-025, Catalog number 6276-1-219, and Catalog number 6276-1-325. Howmedica Osteonics Corp, Stryker Ireland, Carrighwohill Sterile. Intended Use: For primary or revision total hip arthroplasty and severe proximal bone loss. Stems are intended to be used with Howmedica Osteonics femoral heads, unipolar and bipolar components, and acetabular components. Conditions for which product may be appropriate for use: rheumatoid arthritis, correction of deformity, non-inflammatory degenerative joint disease, revision surgery, treatment of non-union, fracture of proximal femur with head involvement.

Recall number
Z-1642-2009
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
Restoration Modular Cone Bodies; titanium Alloy and Hydroxyapatite, Sterile. Catalog number 6276-1-021, Catalog number 6276-1-025, Catalog number 6276-1-219, and Catalog number 6276-1-325. Howmedica Osteonics Corp, Stryker Ireland, Carrighwohill Sterile. Intended Use: For primary or revision total hip arthroplasty and severe proximal bone loss. Stems are intended to be used with Howmedica Osteonics femoral heads, unipolar and bipolar components, and acetabular components. Conditions for which product may be appropriate for use: rheumatoid arthritis, correction of deformity, non-inflammatory degenerative joint disease, revision surgery, treatment of non-union, fracture of proximal femur with head involvement.
Status
Terminated
Initiated
2009-04-02
Posted
2009-05-21
Terminated
2012-08-27
Reason
Packaging burst test values for the Restoration Modular Cone Body family were found to be incorrectly transferred from the finalized validation to the work specification. Testing showed that burst test results for recalled products fell below the validated parameters.
Root cause
Process control
Action
Urgent Product Recall letters were sent via Federal Express on April 2, 2009 to Stryker branches, Chief of Orthopaedics, Hospital Rick Management and Surgeons. The letter stated the issue, the potential hazards, and the risk mitigation. Customers were asked to examine their inventory and hospital locations to identify the affected products; retrieve all effected products and return it to their branch or agency warehouse for reconciliation; reconcile all products on the attached Product Accountability Form and fax a copy to the number listed; return the affected product via the Inter-Org Transfer and ship it to the attention of the individual listed; attach the fluorescent orange PRODUCT REMEDIATION sticker to their return, indicating the Product Remediation #, and mark the box with the words "PRODUCT RECALL". Questions should be directed to Colleen O'Meara at 201-831-5970.
Quantity
17 units
Distribution
Nationwide Distribution
Lot, serial or model codes
Lot # 211080002; Lot # 22109602; Lot # 21883201; Lot # 21910302 and 21883401.
510(k) numbers
K013106, K022549, K040734
PMA numbers
not on file
Product code
LZO
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source