Skip to the content

FDA Medical Device Recalls (openFDA)

Merit Medical Systems, Inc.: Merit Medical, REF: R2007, EsophyX Z+, (1) Fastener Delivery Device, SterileEO, RxOnly

Recall number
Z-1640-2025
Recalling firm
Merit Medical Systems, Inc.
Product
Merit Medical, REF: R2007, EsophyX Z+, (1) Fastener Delivery Device, SterileEO, RxOnly
Status
Open, Classified
Initiated
2025-03-14
Posted
2025-04-21
Terminated
not on file
Reason
Update to IFU to provide additional information related to risk with device over-rotation and multiple deployment of fasteners.
Root cause
Under Investigation by firm
Action
On March 14, 2025, Merit Medical issued a "Urgent Medical Device Recall Notice via E-Mail, followed by mail to affected consignees. Merit asked consignees to take the following actions: 1. Read and understand this communication and the updated IFU referenced in the attached poster. 2. There is no need to return product to Merit. 3. Post the attached poster on or near the affected products so users are aware of the updated IFU. 4. Discard the current IFU with the product at point of use. 5. Ensure that applicable personnel within your organization are made aware of this communication. 6. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them. 7. If you require paper IFUs, please contact RESPONSE@merit.com. 8. Please acknowledging you are aware of the information communicated and have completed the actions required of you. 9. If you require additional product training regarding the IFU updates, please contact your Merit Sales Representative to schedule.
Quantity
1655 units
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV and the countries of Brazil, Egypt, Israel, Saudi Arabia, Spain, Thailand, Turkey, United Arab Emirates.
Lot, serial or model codes
All Lots/ UDI: 00810275011089
510(k) numbers
["K240879"]
PMA numbers
not on file
Product code
ODE
Firm FEI
1721504
Firm city
South Jordan
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register