Baxter Healthcare Corp.: Logix TPN Software Solution, catalog 2M8410, consists of the Logix Order Entry (OE) Installation CD, version 1.0.28 and Logix Compounder Manager (CM) Installation CD, version 3.1.8. The software is included in a number of Baxter's pharmacy compounders. The software assists the pharmacist in management of compounding, creating and authorizing nutritional solutions intended for IV administration.
- Recall number
- Z-1640-2008
- Recalling firm
- Baxter Healthcare Corp.
- Product
- Logix TPN Software Solution, catalog 2M8410, consists of the Logix Order Entry (OE) Installation CD, version 1.0.28 and Logix Compounder Manager (CM) Installation CD, version 3.1.8. The software is included in a number of Baxter's pharmacy compounders. The software assists the pharmacist in management of compounding, creating and authorizing nutritional solutions intended for IV administration.
- Status
- Terminated
- Initiated
- 2008-03-28
- Posted
- 2008-08-24
- Terminated
- 2008-08-25
- Reason
- Software anomaly results in inaccurate information being printed out on the Delivery Report, even though the compounder performed the compounding correctly.
- Root cause
- Software design
- Action
- Baxter sent the 3/28/08 Urgent Device Correction letter to affected customers, to the attention of the Director of Pharmacy, via first class mail to advise them of the potential inaccuracy of the Delivery Report due to a software anomaly. The accounts were instructed to review each delivery report for accuracy. If the report contains zero actual volumes the user must manually reprint the report. The letter outlined the process for reprinting the Delivery Report. Any questions were directed to the Baxter Nutritional Professional Services at 1-800-422-2751. The accounts were requested complete the enclosed reply form acknowledging receipt and understanding of the letter and dissemination of the information to their staff. A software upgrade will be provided. If you have questions about this letter or you did not receive this letter contact Baxter Nutrition Professional Services at 1-800-422-2751.
- Quantity
- 5 units
- Distribution
- Arizona, Minnesota, Wisconsin, Puerto Rico and Canada
- Lot, serial or model codes
- Software version 1.0.2 (comprised of Logix OE version 1.0.28 and Logix CM version 3.1.8)
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LHI
- Firm FEI
- 1416980
- Firm city
- Round Lake
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28