Skip to the content

FDA Medical Device Recalls (openFDA)

Churchill Medical Systems, Inc.: Churchill Medical Systems Bag Access Device w/Smartsite Needleless Injection Site, Sterile, Non pyrogenic, Latex Free Product Code: KSA-140. Churchill Medical Systems, Inc. Dover, NH 03820. Intravascular administration set.

Recall number
Z-1639-2010
Recalling firm
Churchill Medical Systems, Inc.
Product
Churchill Medical Systems Bag Access Device w/Smartsite Needleless Injection Site, Sterile, Non pyrogenic, Latex Free Product Code: KSA-140. Churchill Medical Systems, Inc. Dover, NH 03820. Intravascular administration set.
Status
Terminated
Initiated
2009-10-20
Posted
2010-05-21
Terminated
2013-03-28
Reason
Bond between the spike and needle may leak, break or detach resulting in leakage or spillage of chemotherapeutic agents
Root cause
Process change control
Action
Churchill Medical Systems issued an "Urgent: Product Advisory Customer Notification" dated October 20, 2009 requesting return of affected product and notification to end users. For further information, contact Churchill Medical Systems at 1-603-743-5988 or via fax at 1-603-743-6328.
Quantity
283,900 each
Distribution
Distributed to Florida.
Lot, serial or model codes
Lot Numbers: 903005, 904057, 904060, 904061, 904062, 904063, 905089, 905090, 905091, 906199, 907005, 907006, 907007, 909014, 909015, 909016, 909017, 909018, 909026, 909031, 909032, 909033, 909034, 909035, 909036 and 909037.
510(k) numbers
["K011336"]
PMA numbers
not on file
Product code
FPA
Firm FEI
1000120555
Firm city
Dover
Firm state
NH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register