Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: MAMMOMAT Inspiration. Intended for mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals.

Recall number
Z-1638-2014
Recalling firm
Siemens Medical Solutions USA, Inc
Product
MAMMOMAT Inspiration. Intended for mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals.
Status
Terminated
Initiated
2014-04-25
Posted
2014-05-20
Terminated
2014-12-15
Reason
There is a potential and possible hazard to the user when using the MAMMOMAT Inspiration PC monitor at the control desk, in that the holder of the PC monitor can break causing an unstable monitor to fall causing possible serious injury.
Root cause
Device Design
Action
A safety advisory notice, dated April 25, 2014, was sent to direct accounts to notify them of the issue and provide instructions to avoid the potential risk of injury.
Quantity
5
Distribution
Distributed in the states of GA, FL, NY, NV, and MA.
Lot, serial or model codes
(Material number 10549894) with serial numbers 5000 through 5029.
510(k) numbers
["K122286"]
PMA numbers
not on file
Product code
MUE
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register