Skip to the content

FDA Medical Device Recalls (openFDA)

Bard Peripheral Vascular Inc: BD WAVELINQ 4F EndoAVF System, Arterial Catheter 50 cm, Venous Catheter 43 cm, Sterile EO, Rx only, Mfr. Bard Peripheral Vascular, Inc., Ref W04200, Lot S0053, UDI (01)00801741182754(17)210215(10)S0053 - Product Usage: The WavelinQ 4F EndoAVF System in indicated for the creation of an arteriovenous fistula (AVF) using concomitant ulnar artery and ulnar vein or concomitant radial artery and radial vein in patients with minimum artery and vein diameters of 2.0 mm at the fistula creation site who have chronic kidney disease and need hemodialysis.

Recall number
Z-1637-2020
Recalling firm
Bard Peripheral Vascular Inc
Product
BD WAVELINQ 4F EndoAVF System, Arterial Catheter 50 cm, Venous Catheter 43 cm, Sterile EO, Rx only, Mfr. Bard Peripheral Vascular, Inc., Ref W04200, Lot S0053, UDI (01)00801741182754(17)210215(10)S0053 - Product Usage: The WavelinQ 4F EndoAVF System in indicated for the creation of an arteriovenous fistula (AVF) using concomitant ulnar artery and ulnar vein or concomitant radial artery and radial vein in patients with minimum artery and vein diameters of 2.0 mm at the fistula creation site who have chronic kidney disease and need hemodialysis.
Status
Terminated
Initiated
2019-04-12
Posted
not on file
Terminated
2021-07-28
Reason
The firm has become aware of a potential problem with their venous and arterial magnetic catheters may be at risk of experiencing a magnetic deficiency, which may result in the magnets failing to attract to one another. This may result in procedural delay in obtaining or creating a functional fistula and need for additional contrast to perform the necessary fluoroscopy. This can pose an incremental risk of harm to a diabetic patient's kidney function.
Root cause
Nonconforming Material/Component
Action
On 04/12/2019, the firm sent an "Urgent Medical Device Recall Notification" to its customers to inform them the firm has become aware that the WavelinQ 4FEndoAVF System, Product Code W04200, Lot Number S0053 may be at risk of experiencing a magnetic deficiency between the venous and arterial magnetic catheters, which may result in the magnets failing to attract to one another. The firm is instructing customers to remove any affected inventory from use and provided instructions to its customers to take regarding this product. The firm is requesting its customers to complete the required actions and fill out the Recall and Effectiveness Check Form to confirm the quantities and lot numbers of each recalled products intended to be returned. Email the completed Recall and Effectiveness Check Form to BPV.CustomerSupportCenter@crbard.com or fax it to BPV at 1-800-994-6772. For additional questions, customers are instructed to call the BPV Customer Support Center at 1-800-321-4254 Option #5. On 04/26/2019, a second communication "Urgent Medical Device Recall Notification" letter was sent to each of the US consignees via FedEx with proof of delivery notification. The letter explained the firm's records showing the customer purchased one or more units from the affected lot. The 2nd notice was send to confirm receipt of this recall information. Firm's records indicated they have not received all returned inventory from the customer. The firm's corrective action plan is still being developed. A field action is also in process of being initiated for affected lot of an equivalent product sold outside the US (EverlinQ 4F EndoAVF System). The equivalent product is not approved or sold in the US.
Quantity
150 Units
Distribution
US Nationwide distribution in the states of NY, TN, TX, FL, NC, GA, NH, MO, CA, SC, AL, IL, VA, IN. No US Gov. Consignees No OUS Consignees for the WavelinQ 4F EndoAVF System. However, a field action is in process of being initiated for affected lot of an equivalent product sold outside the US (EverlinQ 4F EndoAVF System). The equivalent product is not approved or sold in the US.
Lot, serial or model codes
Product Code: W04200  Lot/Serial Number: S0053
510(k) numbers
["K182796"]
PMA numbers
not on file
Product code
PQK
Firm FEI
2020394
Firm city
Tempe
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register