Smith & Nephew Inc: TRIGEN INTERTAN 1.5 NAIL: (1) REF 71676515, 10 MM X 10 CM, 125 degree, LEFT, (2) REF 71676556, 11.5 MM X 36 CM, 125 degree, RIGHT, (3) REF 71676558, 11.5 MM X 38 CM, 125 degree, RIGHT, (4) REF 71676561, 11.5 MM X 42 CM, 125 degree, LEFT, Smith & Nephew, Inc. Orthopedic fixation device.

Recall number
Z-1637-2014
Recalling firm
Smith & Nephew Inc
Product
TRIGEN INTERTAN 1.5 NAIL: (1) REF 71676515, 10 MM X 10 CM, 125 degree, LEFT, (2) REF 71676556, 11.5 MM X 36 CM, 125 degree, RIGHT, (3) REF 71676558, 11.5 MM X 38 CM, 125 degree, RIGHT, (4) REF 71676561, 11.5 MM X 42 CM, 125 degree, LEFT, Smith & Nephew, Inc. Orthopedic fixation device.
Status
Terminated
Initiated
2014-04-04
Posted
2014-05-20
Terminated
2017-07-01
Reason
A manufacturing error resulted in the set screw being inserted too deep in the affected nails.
Root cause
Process control
Action
The firm initiated the recall via an "Urgent - Product Recall 1st Notification - Urgent" letter and e-mail on 04/04/2014, identifying the product, explaining the problem, and the potential risk associated with using the product. Customers are to inspect their inventory, and locate and quarantine any affected product. Additionally, the Inventory Return Certification form should be completed and returned along with any affected product identified, once a Return Authorization number has been obtained from Smith & Nephew's Safety Affairs Department.
Quantity
1,253 units total
Distribution
Worldwide Distribution -- USA, France, Germany, Sweden, Switzerland, UK, Canada, Australia, Japan, China, India, Dubai, South Africa, South Korea, Mexico, and Venezuela.
Lot, serial or model codes
(1) REF 71676515, Lot Number 13LT36602  (2) REF 71676556, Lot Numbers: 13LT36604, 13MT37354, 13MT37947, 14AT38705, 14AT38714  (3) REF 71676558, Lot Numbers: 13KT36282, 13MT37356, 13MT37664, 13MT37949, 14AT38187  (4) REF 71676561, Lot Numbers: 13LT36605, 13MT37358, 14AT38188
510(k) numbers
["K040212"]
PMA numbers
not on file
Product code
JDS
Firm FEI
1020279
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register