Skip to the content

FDA Medical Device Recalls (openFDA)

Tandem Diabetes Care, Inc.: Tandem pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion set, into a patient s subcutaneous tissue. As with current insulin infusion pumps on the U.S. market, the desired timing and quantity of the insulin delivery is programmed by the user (i.e., the patient).

Recall number
Z-1636-2026
Recalling firm
Tandem Diabetes Care, Inc.
Product
Tandem pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion set, into a patient s subcutaneous tissue. As with current insulin infusion pumps on the U.S. market, the desired timing and quantity of the insulin delivery is programmed by the user (i.e., the patient).
Status
Open, Classified
Initiated
2026-01-30
Posted
2026-03-25
Terminated
not on file
Reason
The Czech language user guide contained multiple translation errors. The most significant error was that it incorrectly instructs users to verify that the infusion set is connected to the body, when it should instead direct users to confirm that the infusion set is not connected to the body. Following this instruction could result in unintended insulin delivery, potentially leading to a hypoglycemic event.
Root cause
Process change control
Action
On January 30, 2026, Tandem Diabetes Care issued a Urgent Medical Device Recall Notice to affected consignees via E-Mail. Tandem ask consignees to take the following actions: 1. Discard your existing User Guide. 2. Refer to the corrected User Guide version 7.8.1 which is available online at www.tandemdiabetes.com/cs-cz/support/resources/documents for accurate instructions on using your t:slim X2 Insulin Pump. 3. Physical copies of the corrected User Guide are being produced to replace the incorrect versions. If you would like a physical copy of the corrected User Guide, you may contact A. Import s customer service line to arrange for a replacement. 4.As requested above, please complete the online form acknowledging receipt of this notice. 5. Should you experience any device-related serious incidents, please report them to your distributor, a local contact point, or the State Institute for Drug Control.
Quantity
1,380 units
Distribution
International distribution in the countries of Czech Republic, Slovakia.
Lot, serial or model codes
Version 7.8.1/ UDI: (01)008500066132625
510(k) numbers
["K232380","K232382","K243823","K250792","K250798"]
PMA numbers
not on file
Product code
QFG
Firm FEI
3007981285
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register