Skip to the content

FDA Medical Device Recalls (openFDA)

Qiagen Sciences LLC: AmniSure ROM Test, Material Nos. FMRT-1-25-US (US Distribution, 25 tests/box), FMRT-1-10-US (US Distribution, 10 tests/box), FMRT-1-25-CA (Canadian Distribution; 25 tests/box)

Recall number
Z-1636-2020
Recalling firm
Qiagen Sciences LLC
Product
AmniSure ROM Test, Material Nos. FMRT-1-25-US (US Distribution, 25 tests/box), FMRT-1-10-US (US Distribution, 10 tests/box), FMRT-1-25-CA (Canadian Distribution; 25 tests/box)
Status
Terminated
Initiated
2020-02-21
Posted
not on file
Terminated
2020-10-05
Reason
An image on the associated polybag contains an incorrect rinse time of 30 (thirty seconds). The correct rinse time is 1 (one) minute.
Root cause
Employee error
Action
On February 21, 2020, the firm distributed Urgent Medical Device Correction letters to affected customers. Customers were informed of the incorrect rinse time on the product's labeling, and reminded of the correct rinse time of 1 (one) minute. Customers were asked to immediately share the contents of the customer letter with all users of the AmniSure ROM Test in their facility to ensure awareness. Test users should refer to the instructions for use (IFU) included in each kit box. The firm is not asking for product to be returned. Customers were asked to complete and return an attached response form.
Quantity
1654 kit boxes
Distribution
Nationwide domestic distribution. Foreign distribution to Canada.
Lot, serial or model codes
FMRT-1-25-US: Lots 563010976, 563011620, 563011164, 5660397, 5660552  FMRT-1-10-US: Lot 563011618  FMRT-1-25-CA: Lot 563011195
510(k) numbers
["K081767"]
PMA numbers
not on file
Product code
NQM
Firm FEI
3003572099
Firm city
Germantown
Firm state
MD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register