Skip to the content

FDA Medical Device Recalls (openFDA)

Hummingbird Med: ABG-HM-1 Hummi Micro Draw Blood Transfer Device Product Usage: It is used for blood transfer and collection from Peripheral Arterial Line Catheters.

Recall number
Z-1636-2016
Recalling firm
Hummingbird Med
Product
ABG-HM-1 Hummi Micro Draw Blood Transfer Device Product Usage: It is used for blood transfer and collection from Peripheral Arterial Line Catheters.
Status
Terminated
Initiated
2015-10-01
Posted
2016-05-20
Terminated
2017-01-26
Reason
Hummingbird Med Devices, Inc. is recalling ABG-HM-1 Hummi Micro Draw Blood Transfer because the connection between the "Hummi" Y connector and the yellow cannula hub may not be sufficiently secure enough and could separate while in use.
Root cause
Component design/selection
Action
On 10/01/15 the firm sent out customer notification letters. The letter states for the following actions to be taken: 1. Customers using the Hummi Micro Draw device are recommended not to use the identified lot. 2. A local distributor/sales representative will be contacting customers to arrange a return of the identified lot for credit and arrangement of replacement. 3. Replacement product should be available within 10 days of 10/30/15. 4. During the period of product unavailability customers might want to choose to temporarily to return their previous method for blood drawing. The firm requests a Reply Form be completed and returned via fax at 949-583-2775 or email to JorgeHaider@hummingbirdmed.com. On 11/19/2015 the firm sent out a second notification letter involving additional recall lots. The firm provided the same instructions to customers as their previous notification; however stated replacement product should be available by mid December 2015.
Quantity
11500 units
Distribution
Distributed in the states of CA, KY, MD, GA, and IL.
Lot, serial or model codes
Lot 15180, 15286, 15287, 15300, 15305
510(k) numbers
not on file
PMA numbers
not on file
Product code
KST
Firm FEI
3009382876
Firm city
Lake Forest
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register