Skip to the content

FDA Medical Device Recalls (openFDA)

Microline Surgical: Microline Renew Fenestrated Control Tip Grasper, Single Patient Use, Product Number: 3222 The ReNew Fenestrated Grasper Forceps Tips are intended to cut, grasp, and dissect various abdominal tissue for use in endoscopic, including laproscopic surgical procedures where instruments are introduced onto the body through a cannula.

Recall number
Z-1635-2014
Recalling firm
Microline Surgical
Product
Microline Renew Fenestrated Control Tip Grasper, Single Patient Use, Product Number: 3222 The ReNew Fenestrated Grasper Forceps Tips are intended to cut, grasp, and dissect various abdominal tissue for use in endoscopic, including laproscopic surgical procedures where instruments are introduced onto the body through a cannula.
Status
Terminated
Initiated
2014-04-14
Posted
2014-05-20
Terminated
2015-09-02
Reason
Grasper jaw may break when force is applied to the jaw
Root cause
Nonconforming Material/Component
Action
Microline Surgical sent an Recall Notice dated April 14, 2014, via overnight mail to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Immediately examine your inventory and quarantine product subject to this recall. In addition, if you have further distributed this product, please identify your customers and notify them at once of this product recall. Your notification to your customers may be enhanced by including a copy of this recall notification letter. Please contact Microline Customer Service ( 978) 922-9810 for a RGA number to return all unused product. This recall should be carried out to the hospital level. Please complete and return the enclosed response form as soon as possible.
Quantity
540 units
Distribution
Worldwide Distribution - US including the states of TX, IL, TN and NY, and the country of Chile.
Lot, serial or model codes
Lot Numbers: 00116169 and 00115866
510(k) numbers
["K981389"]
PMA numbers
not on file
Product code
GEI
Firm FEI
1000137656
Firm city
Beverly
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register