Skip to the content

FDA Medical Device Recalls (openFDA)

Smith & Nephew, Inc.: NAVIO Soft Tissue Protector, included in instrument kits. The NAVIO Tissue Protector is a reusable instrument designed to, when used as instructed, provide the required bone pin spacing for the Navio bone clamps and potentially reduce the amount of patient tissue affected when placing the bone pins.

Recall number
Z-1634-2020
Recalling firm
Smith & Nephew, Inc.
Product
NAVIO Soft Tissue Protector, included in instrument kits. The NAVIO Tissue Protector is a reusable instrument designed to, when used as instructed, provide the required bone pin spacing for the Navio bone clamps and potentially reduce the amount of patient tissue affected when placing the bone pins.
Status
Terminated
Initiated
2020-02-27
Posted
not on file
Terminated
2024-08-02
Reason
Potential for the NAVIO Soft Tissue Protector to become stuck or bound to the bone pin intraoperatively when inserting the NAVIO bone pins.
Root cause
Device Design
Action
The firm, Smith & Nephew, notified customers on about 03/03/2020 via "Urgent Medical Device Recall Notice" letter. The customers were Instructed to examine inventory and quarantine affected devices, notify customers if further distributed, and complete and return the Response Form to fieldactions@smith-nephew.com or fax to 1-901-566-7975. Questions and concerns can be directed to EieldActions@smlth-nephew .com. or 901-399-1970.
Quantity
348 units
Distribution
Nationwide distribution to AL, AZ, CA, CT, FL, GA, HI, IA, IL, KS, LA, MA, MI, MN, MS, MT, NC, NJ, NY, OH, OK, PA, TN, TX, VA, WA, WI, WV. International distribution to Turkey, United Kingdom, Thailand, Switzerland, France, Australia.
Lot, serial or model codes
Component Product Number: PFSR101092;   a. Instrument Kit Number: PFSR02050, Lot Numbers 2067, 2212, 3809, 3869, 4354, 5228, 5337, 5494, 5514, 5887, 6248;   b. Instrument Kit Number: PFSR02051, Lot Numbers 4354, 5514;   c. Instrument Kit Number: PFSR02052, Lot Numbers 4354, 5494, 6248;   d. Instrument Kit Number: PFSR02053, Lot Numbers 6248
510(k) numbers
["K121936"]
PMA numbers
not on file
Product code
OLO
Firm FEI
1020279
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register