Leica Microsystems, Inc.: Leica Wireless Footswitch, Type B used in conjunction with Leica Surgical Microscopes - Models M620, M844/820, M525 and M720; There are two wireless models: a) MFS20, stock number 10448406, a 12 function wireless footswitch type B with cross pedals, XY-actuator 10717602 and screw 10630134 for dust protection enclosed; b) MFS21, stock number 10448407, a 16 function wireless footswitch type B with cross pedals, XY-actuator 10717602 and screw 10630134 for dust protection enclosed;

Recall number
Z-1633-2010
Recalling firm
Leica Microsystems, Inc.
Product
Leica Wireless Footswitch, Type B used in conjunction with Leica Surgical Microscopes - Models M620, M844/820, M525 and M720; There are two wireless models: a) MFS20, stock number 10448406, a 12 function wireless footswitch type B with cross pedals, XY-actuator 10717602 and screw 10630134 for dust protection enclosed; b) MFS21, stock number 10448407, a 16 function wireless footswitch type B with cross pedals, XY-actuator 10717602 and screw 10630134 for dust protection enclosed;
Status
Terminated
Initiated
2010-03-29
Posted
2010-05-20
Terminated
2011-02-15
Reason
The pedals of the Footswitch Type B stick and the Zoom/Focus moves to the end position.
Root cause
Other
Action
Leica Microsystems sent Medical Device Correction letters dated 3/29/10 to the affected customers via first class mail on the same date, informing them of the potential for injury to the patient's eye if the cross pedal of the footswitch sticks. They were informed that their local Leica Microsystems representative will contact them to arrange for the replacement of the cross pedal. The letters also provided the interim necessary steps required to avoid the risk of patient injury while they are awaiting the cross pedal replacement, i.e. do not use the affected footswitches with a surgical microscope that has a BIOM accessory. The accounts were requested to complete the attached Acknowledgement Form, listing the serial numbers of the units at their facility and fax it back to Leica at 1-847-607-3147. A company contact was provided in the letter for questions.
Quantity
3 units
Distribution
Worldwide distribution -- U.S. (nationwide), Canada, Australia, Europe and Asia
Lot, serial or model codes
stock number 10448406, serial numbers 60110005, 60110004, 60110003, 60110002, 60110001, 50110001 and 120110001;  stock number 10448407, serial numbers 80110003 and 221209001
510(k) numbers
not on file
PMA numbers
not on file
Product code
FSO
Firm FEI
3000210133
Firm city
Bannockburn
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register