Siemens Healthcare Diagnostics, Inc.: Atellica CH Toxicology Calibrator (TOX CAL)-The Atellica CH Toxicology Calibrator (TOX CAL) is for in vitro diagnostic use in calibrating the Acet, ETOH and Sal assays using the Atellica CH Analyzer Siemens Material Number (SMN): 11099440

Recall number
Z-1632-2023
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
Atellica CH Toxicology Calibrator (TOX CAL)-The Atellica CH Toxicology Calibrator (TOX CAL) is for in vitro diagnostic use in calibrating the Acet, ETOH and Sal assays using the Atellica CH Analyzer Siemens Material Number (SMN): 11099440
Status
Open, Classified
Initiated
2023-04-05
Posted
2023-05-25
Terminated
not on file
Reason
Reassignment of the Atellica CH Toxicology Calibrators (TOX CAL) and ADVIA Chemistry ToxAmmonia Calibrators for the Atellica CH Salicylate (Sal) Assay due to a positive bias
Root cause
Under Investigation by firm
Action
Siemens issued Urgent Medical Device Corrections (UMDC) ACHC23-03.A.US and ACHC23- 03.A.US.CHC to US customers via FedEx beginning on 04/05/2023. An Urgent Field Safety Notices (UFSN) ACHC23-03.A.OUS and ACHC23-03.A.OUS.CHC issued to OUS Siemens reporting sites with affected customers, for distribution and regional reporting to their local competent authorities beginning on 04/05/2023. Letter states reason for recall, health risk and action to take: Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Until the updated lot-specific value sheets are available on Document Library, keep a copy of this letter as a reference for the updated salicylate calibrator values. Perform the instructions provided in Additional Information section below. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
Quantity
6546 units US; 3681 units OUS
Distribution
US Nationwide and Worldwide Distribution Foreign: Argentina Australia Austria¿¿¿¿¿¿¿¿ Bahrain¿¿¿¿¿¿¿¿ Bangladesh Belgium¿¿¿¿¿¿¿¿ Bosnia¿Herzeg.¿ Brazil Canada Chile China Colombia Czech¿Republic¿ Egypt¿¿¿¿¿¿¿¿¿¿ Estonia¿¿¿¿¿¿¿¿ Finland¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Georgia¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Hong Kong Hungary¿¿¿¿¿¿¿¿ India Israel¿¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Latvia¿¿¿¿¿¿¿¿¿ Malaysia Mexico Netherlands¿¿¿¿ Norway¿¿¿¿¿¿¿¿¿ Pakistan¿¿¿¿¿¿¿ Poland¿¿¿¿¿¿¿¿¿ Portugal¿¿¿¿¿¿¿ Republic of Korea Saudi¿Arabia¿¿¿ Serbia¿¿¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ South Africa Spain¿¿¿¿¿¿¿¿¿¿ Sweden¿¿¿¿¿¿¿¿¿ Taiwan Turkey¿¿¿¿¿¿¿¿¿ United¿Kingdom¿ Uruguay Vietnam
Lot, serial or model codes
UDi-DI: 00630414597614 Lot Numbers: 74285306 Exp. Date: 2023-05-31; 74536672 Exp Date: 2024-03-31; 74732386 Exp. Date: 2024-11-30
510(k) numbers
not on file
PMA numbers
not on file
Product code
JLX
Firm FEI
2432235
Firm city
Tarrytown
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register