Skip to the content

FDA Medical Device Recalls (openFDA)

Surgical Instrument Service And Savings, Inc.: Barron Radial Vacuum Trephine (1) Complete Trephine Set 9.0mm Dia. Ref. K20-2062 Lot 122336 Opthalmic

Recall number
Z-1632-2013
Recalling firm
Surgical Instrument Service And Savings, Inc.
Product
Barron Radial Vacuum Trephine (1) Complete Trephine Set 9.0mm Dia. Ref. K20-2062 Lot 122336 Opthalmic
Status
Terminated
Initiated
2013-05-03
Posted
2013-07-01
Terminated
2013-07-11
Reason
The seal which maintains a sterile barrier for reprocessed medical devices became compromised to the point where product may fall out of the pouch.
Root cause
Packaging process control
Action
MEDISISS sent a Correction Immediate Action Request letter dated May 2, 2013, to all affected customes. The letter identified the product, the problem, and the action to be taken by the customer. The letter directed the customer to inspect the products prior to use for seal integrity as specified on the product labeling: "Sterile unless package has been damaged or opened. Do not use if package is damaged." This will be followed up by a visit from the customer's sales representatives who are tasked with inspecting all the lots that may be remaining in the customer's inventory. The sales representative is directed to complete a "Verification of Seal Inspection" form and return it to MEDISISS Customer Relations. Devices with unacceptable seals are to be returned to MEDISISS. The sales representative is responsible for obtaining a shipping label from customer service to enable the devices to be returned to the recalling firm via Fed Ex. Customers with questions were instructed to call 866-866-7477.
Quantity
1
Distribution
Nationwide Distribution.
Lot, serial or model codes
122336
510(k) numbers
not on file
PMA numbers
not on file
Product code
HRH
Firm FEI
3002299608
Firm city
Redmond
Firm state
OR
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register