FDA Medical Device Recalls (openFDA)
Leica Microsystems, Inc.: Leica Footswitch, Type B used in conjunction with Leica Surgical Microscopes - Models M620, M844/820, M525 and M720; There are three cable connection footswitch models : a) MFS11, stock number 10445599, a 6 function footswitch type B with cross pedals; b) MFS12, stock number 10445600, a 12 function footswitch type B with cross pedals, XY-actuator 10717602 and screw 10630134 for dust protection enclosed; c) MFS18, stock number 10448164, a 16 function footswitch type B with cross pedals, XY-actuator 10717602 and screw 10630134 for dust protection enclosed;
- Recall number
- Z-1632-2010
- Recalling firm
- Leica Microsystems, Inc.
- Product
- Leica Footswitch, Type B used in conjunction with Leica Surgical Microscopes - Models M620, M844/820, M525 and M720; There are three cable connection footswitch models : a) MFS11, stock number 10445599, a 6 function footswitch type B with cross pedals; b) MFS12, stock number 10445600, a 12 function footswitch type B with cross pedals, XY-actuator 10717602 and screw 10630134 for dust protection enclosed; c) MFS18, stock number 10448164, a 16 function footswitch type B with cross pedals, XY-actuator 10717602 and screw 10630134 for dust protection enclosed;
- Status
- Terminated
- Initiated
- 2010-03-29
- Posted
- 2010-05-20
- Terminated
- 2011-02-15
- Reason
- The pedals of the Footswitch Type B stick and the Zoom/Focus moves to the end position.
- Root cause
- Other
- Action
- Leica Microsystems sent Medical Device Correction letters dated 3/29/10 to the affected customers via first class mail on the same date, informing them of the potential for injury to the patient's eye if the cross pedal of the footswitch sticks. They were informed that their local Leica Microsystems representative will contact them to arrange for the replacement of the cross pedal. The letters also provided the interim necessary steps required to avoid the risk of patient injury while they are awaiting the cross pedal replacement, i.e. do not use the affected footswitches with a surgical microscope that has a BIOM accessory. The accounts were requested to complete the attached Acknowledgement Form, listing the serial numbers of the units at their facility and fax it back to Leica at 1-847-607-3147. A company contact was provided in the letter for questions.
- Quantity
- 54 units (52 - U.S., 2 - Canada)
- Distribution
- Worldwide distribution -- U.S. (nationwide), Canada, Australia, Europe and Asia
- Lot, serial or model codes
- stock number 10445599, serial numbers 90909001 and 301109002; stock number 10445600, serial numbers 91009001, 81209001, 80909001, 8011001, 71009002, 71009001, 70110002, 61009002, 41209004, 41209002, 291009004, 271009004, 261109001, 250909001, 230909001, 211209005, 211209004, 211009004, 180909002, 180909001, 180110002, 171209002, 161209002, 160909001, 151209001, 140110001, 121009001, 120110002, 120110001, 111209002, 111209001, 110909003, 110909002, 110909001, 11009003, 101209003, 101209002, 101109002 and 100909001; stock number 10448164, serial numbers 80909001, 71209001, 70909002, 31209004, 31209003, 31209002, 31209001, 301009001, 300909001, 291009001, 290909002, 290909001, 240909002, 240909001, 230909001, 180909005, 180909004, 160909002, 100909002 and 100909001;
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FSO
- Firm FEI
- 3000210133
- Firm city
- Bannockburn
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28