Skip to the content

FDA Medical Device Recalls (openFDA)

Koven Technology, Inc.: Brand Name: BiDop 3 Product Name: BiDop 3 Pocket Doppler (ES-100V3) with BT2M20S8C fetal probe Model/Catalog Number: ES-100V3 (device); BT2M20S8C (2 MHz fetal probe). Software Version: N/A Product Description: Hand-held, battery-powered ultrasonic Doppler blood-flow detector with LCD numeric display, used with interchangeable probes to detect vascular blood-flow signals and display heart rate; affected units were packaged with a 2 MHz BT2M20S8C fetal probe and labeled for both vascular and fetal applications.

Recall number
Z-1631-2026
Recalling firm
Koven Technology, Inc.
Product
Brand Name: BiDop 3 Product Name: BiDop 3 Pocket Doppler (ES-100V3) with BT2M20S8C fetal probe Model/Catalog Number: ES-100V3 (device); BT2M20S8C (2 MHz fetal probe). Software Version: N/A Product Description: Hand-held, battery-powered ultrasonic Doppler blood-flow detector with LCD numeric display, used with interchangeable probes to detect vascular blood-flow signals and display heart rate; affected units were packaged with a 2 MHz BT2M20S8C fetal probe and labeled for both vascular and fetal applications.
Status
Open, Classified
Initiated
2026-03-04
Posted
2026-03-24
Terminated
not on file
Reason
Product labeling includes a fetal indication for use that is not cleared under its 510(k).
Root cause
Labeling Change Control
Action
On March 4, 2026, firm began notifying customers via first-class mail. Customers were informed they may still use the BiDop 3 device for vascular applications, but that the device is not cleared for fetal usage. Koven Technology, Inc. will provide a replacement operation manual for devices that are being amended to remove fetal use indications.
Quantity
34
Distribution
US Nationwide distribution in the states of LA, NC, MT, TX MA, OR, NY, IL, CO, GA, FL, WA, WI, MO, AZ, IN.
Lot, serial or model codes
Lot Code: BIDOP 3 #00894912002043 BT2M20S8C #00894912002326
510(k) numbers
["K954397"]
PMA numbers
not on file
Product code
DPW
Firm FEI
1937397
Firm city
Saint Louis
Firm state
MO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register