Skip to the content

FDA Medical Device Recalls (openFDA)

Randox Laboratories Ltd.: Lipase, Catalog number: LI3837; GTIN: 05055273204230 and Catalog number: LI8361; GTIN: 05055273214284 Product Usage: For Professional Use for the quantitative in vitro determination of Lipase in human serum and plasma. Lipase measurements are used in the diagnosis and treatment of diseases of the pancreas such as acute pancreatitis and obstruction of the pancreatic duct.

Recall number
Z-1631-2018
Recalling firm
Randox Laboratories Ltd.
Product
Lipase, Catalog number: LI3837; GTIN: 05055273204230 and Catalog number: LI8361; GTIN: 05055273214284 Product Usage: For Professional Use for the quantitative in vitro determination of Lipase in human serum and plasma. Lipase measurements are used in the diagnosis and treatment of diseases of the pancreas such as acute pancreatitis and obstruction of the pancreatic duct.
Status
Terminated
Initiated
2018-03-02
Posted
not on file
Terminated
2019-04-11
Reason
Randox has now released further steps to avoid contamination of the Lipase Assay on Rx instruments.
Root cause
Under Investigation by firm
Action
On March 1, 2018 Randox Laboratories Ltd. sent a Urgent Medical Device Correction Notification to affected customers. These customers were instructed to: 1) Update the user manual for the RX Instrument with the attached technical bulletin; 2) Replace the Instructions for Use contained within the kit with the revised version; 3) Discuss the contents of this notice with your Medical Director; 4) Complete and return the vigilance response section of this form to technical.services@randox..com within five working days; 5) Send a copy of the FSN to all affected customers and to those who need to be aware within your organization. Customer with questions may contact Randox Technical Services via e-mail technical.services@randox.com or by phone +44 (0) 28 9445 1070
Quantity
8
Distribution
US Nationwide Distribution in the states of West Virginia and Puerto Rico
Lot, serial or model codes
Lot numbers: 428000, 404175, and 404245
510(k) numbers
not on file
PMA numbers
not on file
Product code
CDT
Firm FEI
1000361607
Firm city
Crumlin
Firm state
North
Firm country
Ireland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register