FDA Medical Device Recalls (openFDA)

Philips Respironics, Inc.: BiPAP A40, Brazil: Continuous ventilator, non-life-supporting. Material Number: BR1111169

Recall number
Z-1630-2021
Recalling firm
Philips Respironics, Inc.
Product
BiPAP A40, Brazil: Continuous ventilator, non-life-supporting. Material Number: BR1111169
Status
Terminated
Initiated
2021-04-06
Posted
not on file
Terminated
2024-01-17
Reason
Certain devices were built with motor assemblies that could contain non-conforming plastic material. This could result in 1. Exposure to off-gassing of Volatile Organic Compounds not normally present creating a potential biosafety or toxicological hazard and 2. Sudden failure of the device causing a Ventilator Inoperative condition with the potential for asphyxia if not immediately identified and addressed by the care provider.
Root cause
Nonconforming Material/Component
Action
In April, 2021, Philips issued an Urgent Medical Device Removal notice to customers via letter notifying them that Philips determined that certain devices were built with motor assemblies that could contain non-conforming plastic material. ACTION TO BE TAKEN BY CUSTOMER / USER 1. Locate all impacted devices listed above immediately STOP USE. All impacted devices must be removed from service. Replace with a non-impacted device if it is being used by a patient. 2. This notice needs to be passed on to all those who must be aware within your organization or to any organization where the impacted devices have been transferred. 3. Once impacted devices have been removed from service and isolated notify your Philips Representative that the impacted devices are ready for return. Your Philips Representative will arrange for return of the device. Replacements will be provided free of charge. 4. Fill out the Business Reply Form that has been provided with this letter and return it to your Philips Representative. This form serves as official acknowledgement that you have fully performed your obligations to complete this medical device recall.
Quantity
4 units
Distribution
Distribution US nationwide to states of: LA, PA, WI, CO, WV, WA, CA, TX, MO, OK, FL, and TN. Brazil, China, and Japan
Lot, serial or model codes
Serial Number: V306120599B37, V3061207124CF, V30612105753F, V306121267414
510(k) numbers
not on file
PMA numbers
not on file
Product code
MNT
Firm FEI
2518422
Firm city
Murrysville
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register