Skip to the content

FDA Medical Device Recalls (openFDA)

CooperSurgical, Inc.: Carter-Thomason II Port Closure System; Model Number(s) CTI-1015P, for use in endoscopic and laparoscopic surgery.

Recall number
Z-1630-2015
Recalling firm
CooperSurgical, Inc.
Product
Carter-Thomason II Port Closure System; Model Number(s) CTI-1015P, for use in endoscopic and laparoscopic surgery.
Status
Terminated
Initiated
2015-03-20
Posted
2015-05-19
Terminated
2015-07-29
Reason
The pad printing around the suture entry holes on the 15mm Suture Guide included in the kit may not have been appropriately manufactured resulting in the pad printing flaking off around the entry holes.
Root cause
Component design/selection
Action
Consignees were notified of the recall by letter sent via FedEx dated March 20, 2015. The letter instructs consignees to do the following: discontinue use of the affected product, complete the attached acknowledgement and receipt form to return the product and schedule a replacement.
Quantity
430 units
Distribution
US Nationwide distribution.
Lot, serial or model codes
Lot 174422
510(k) numbers
["K133414"]
PMA numbers
not on file
Product code
NLM
Firm FEI
1216677
Firm city
Trumbull
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register