Skip to the content

FDA Medical Device Recalls (openFDA)

SCM True Air Technologies LLC: JK Bariatric Beds, True Air Technologies, Inc.

Recall number
Z-1630-2014
Recalling firm
SCM True Air Technologies LLC
Product
JK Bariatric Beds, True Air Technologies, Inc.
Status
Terminated
Initiated
2014-03-18
Posted
2014-05-19
Terminated
2018-06-11
Reason
The firm is recalling the products due to lack of design controls, lack of a Device Master Record, and failure to approve specific vendors.
Root cause
Nonconforming Material/Component
Action
SCM True Air Technologies LLC sent an Urgent Medical Device Recall letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to remove the affected product from use and return to True Air Technologies immediately. The units should be decontaminated and bagged with a certification of disinfection before return to True Air Technologies. True Air will provide replacement units to replace the returned units. Customers were instructed to refer to the Medical Device Recall Return Response Form attached for acknowledgement of receipt and contact information for this voluntary recall. Customers with questions should call 800-682-7163.
Quantity
24 beds
Distribution
Nationwide Distribution including GA, KY, and MI.
Lot, serial or model codes
Serial #'s 41160, 10127, 09504, 41918, 106800, 62188, 106805, 42331, 106751, 00234A, 62422, 7987, 50147, 7111, 11151100, BB111, 00284A, 2171269, 62383, 4731, 4732, 4733, 4734, 4735
510(k) numbers
not on file
PMA numbers
not on file
Product code
OSI
Firm FEI
3001452084
Firm city
Louisville
Firm state
KY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register