Skip to the content

FDA Medical Device Recalls (openFDA)

Orthalign, Inc: Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 1 -5 REF:402566 component Lantern Knee System

Recall number
Z-1629-2026
Recalling firm
Orthalign, Inc
Product
Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 1 -5 REF:402566 component Lantern Knee System
Status
Open, Classified
Initiated
2026-02-10
Posted
2026-03-24
Terminated
not on file
Reason
Drill plates manufactured out of specification can result in a bone cut in excess 00800"(2.8448mm)
Root cause
Nonconforming Material/Component
Action
On February 10, 2026, OrthAlign issued a Urgent: Medical device Recall Notification via overnight mail. OrthAlign asked consignees to take the following actions: Please review your inventory for the affected products listed in Table 1. 1. Immediately remove and quarantine the affected products until the products are returned to OrthAlign. 2. Please share this recall notification with all members in your network as well as those where the affected devices have been transferred. 3. If you have further distributed the affected products, please identify and notify consignees. You may provide consignees a copy of this notification letter. 4. Please complete the attached response form within Five (5) business days and email the completed form to : oaregulatory@orthalign.com 5. Adverse actions are quality issues and can be reported to the Food and Drug Administration s MedWatch program: o Online: By completing and submitting the report online at www.fda.gov/medwatch o Regular mail or fax): download the form from www.fda.gov/MedWatch/getforms.htm or call 800-332-1088 to request a reporting form; then complete the form and mail it to the address on the pre-addressed form or submit by fax to 800-332-0178.
Quantity
7 units
Distribution
US Nationwide distribution in the state of UT.
Lot, serial or model codes
Lot # 25091103BB/ UDI: 00810832032427
510(k) numbers
["K200892"]
PMA numbers
not on file
Product code
OLO
Firm FEI
3007521480
Firm city
Aliso Viejo
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register