Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Corporation Englewood: ABACUS SE (Single-Workstation Edition) and ABACUS ME (Multi-Workstation Edition) Product Usage: The ABACUS Software is a Windows-based order entry software application for comprehensive total parenteral nutrition (TPN) calculations and label printing.

Recall number
Z-1628-2015
Recalling firm
Baxter Corporation Englewood
Product
ABACUS SE (Single-Workstation Edition) and ABACUS ME (Multi-Workstation Edition) Product Usage: The ABACUS Software is a Windows-based order entry software application for comprehensive total parenteral nutrition (TPN) calculations and label printing.
Status
Terminated
Initiated
2015-04-08
Posted
2015-05-18
Terminated
2016-12-01
Reason
Baxter Corporation is conducting a field action for the ABACUS SE and ME models due to the possibility that the compounder will load the incorrect formula upon scanning the bag label when two or more different formula files have been created with the same order number.
Root cause
Software design
Action
Baxter sent a Safety Alert letter dated April 8, 2015 to affected customers via USPS first class mail. The letter identified the affected product, problem and actions to be taken. Customers were asked to 1) contact Baxter Technical Services if you need to restore the ABACUS databse, 2) foword a copy of this Safety Alert letter to other facilities or departments within your institution to ensure that those locations are awae of this action, and 3) complete the customer reply form and return it to Baxter by faxing to 224-270-5457 or scanning and e-mailing it to fca@baxter.com. For questions contact the Center for One Baxter at 800-422-9837.
Quantity
882
Distribution
Worldwide Distribution - US Nationwide
Lot, serial or model codes
Product Codes: 8300-0168 ,8300-0169 ,8300-0157 ,8300-0158 ,8300-0046 ,8300-0047 ,8300-0046 ,8300-0047 ,8300-0046 ,8300-0047 ,8300-0046 ,8300-0047,
510(k) numbers
not on file
PMA numbers
not on file
Product code
NEP
Firm FEI
1419106
Firm city
Englewood
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register