Skip to the content

FDA Medical Device Recalls (openFDA)

Arcadia Medical Corporation: Silicone Cuffed Wire Reinforced Extended Connect Pediatric Perfect Fit Neck Flange Tracheostomy Tube, Size 4.0 mm I.D. Catalog Number 353-040. Intended for use in providing direct tracheal access for airway management.

Recall number
Z-1628-2012
Recalling firm
Arcadia Medical Corporation
Product
Silicone Cuffed Wire Reinforced Extended Connect Pediatric Perfect Fit Neck Flange Tracheostomy Tube, Size 4.0 mm I.D. Catalog Number 353-040. Intended for use in providing direct tracheal access for airway management.
Status
Terminated
Initiated
2012-04-18
Posted
2012-05-21
Terminated
2012-08-13
Reason
A defect in 3 tracheostomy tubes. The defect observed was a separation in the bond between the proximal junction of the neck flange and the cannula.
Root cause
Process control
Action
Arcadia notified the 3 consignees with a MEDICAL DEVICE VOLUNTARY RECALL notification dated 4/18/12. The notification identified the affected product, along with the reason for the recall. Consignees were instructed to cease distribution/use of any unopened product that they have remaining and return to firm following the steps provided. Arcadia will replace the product at no charge to the customer. Consignees were to confirm receipt of the recall notification by emailing Trena Depel, trena@arcadiamedical.com. Phone: 925-918-0595 Fax: (415) 762-5489
Quantity
01103307 (50 units) and 01107055 (73 units)
Distribution
Worldwide Distribution - US, including Mariemount, OH and countries of UK and France.
Lot, serial or model codes
Lot numbers: 01103307 (50 units) and 01107055 (73 units)
510(k) numbers
["K031553"]
PMA numbers
not on file
Product code
BTO
Firm FEI
3004590970
Firm city
Crown Point
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register