Ethicon Endo-Surgery Inc: The ECHELON 60 Endoscopic Linear Cutter - Straight, compact ECHELON 60 Endoscopic Linear Cutter - Straight, and the Long ECHELON 60 Endoscopic Linear Cutter - Straight are sterile, single patient use instruments that deliver staples while simultaneously dividing tissue between rows. ECHELON 60 instruments with white, blue, gold, and green reloads deliver two triple-staggered rows of staples. The instruments have a staple line that is approximately 60 mm long and a cut line that is approximately 57 mm long. A staple retaining cap on the reload protects the staple leg points during shipping and transportation. Sterile products are individually packed in a transparent tray sealed with Tyvek. Each tray is placed in an individual carton. 3 individual cartons are placed in a corrugated shipper. The ECHELON 60 Endoscopic Linear Cutter Straight, Compact ECHELON 60 Endoscopic Linear Cutter Straight, and the Long ECHELON 60 Endoscopic Linear Cutter Straight are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue buttressing materials.

Recall number
Z-1627-2012
Recalling firm
Ethicon Endo-Surgery Inc
Product
The ECHELON 60 Endoscopic Linear Cutter - Straight, compact ECHELON 60 Endoscopic Linear Cutter - Straight, and the Long ECHELON 60 Endoscopic Linear Cutter - Straight are sterile, single patient use instruments that deliver staples while simultaneously dividing tissue between rows. ECHELON 60 instruments with white, blue, gold, and green reloads deliver two triple-staggered rows of staples. The instruments have a staple line that is approximately 60 mm long and a cut line that is approximately 57 mm long. A staple retaining cap on the reload protects the staple leg points during shipping and transportation. Sterile products are individually packed in a transparent tray sealed with Tyvek. Each tray is placed in an individual carton. 3 individual cartons are placed in a corrugated shipper. The ECHELON 60 Endoscopic Linear Cutter Straight, Compact ECHELON 60 Endoscopic Linear Cutter Straight, and the Long ECHELON 60 Endoscopic Linear Cutter Straight are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue buttressing materials.
Status
Terminated
Initiated
2012-03-02
Posted
2012-05-18
Terminated
2013-03-22
Reason
There is a possibility of a component breakage, which may cause the device to stay locked on tissue after firing. If the component exhibits this issue, the device does not open either with the anvil release button or with the manual firing release lever.
Root cause
Vendor change control
Action
Ethicon Endo-Surgery sent an Urgent Device Recall letter dated March 19, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to examine their inventory immediately to determine if they have the affected product and to remove the affected product. Fill out the Business Reply Form and fax it to 1-513-337-4138 within 3 business days, even if they do not have the affected product in their inventory. Ethicon will automatically replace any affected product returned by June 29, 2012. Customers were asked to share this information with appropriate staff. If customers should need clinical or product support call 1-800-873-3636, option 6, or contact their local sales representative. For questions regarding this recall call 513-337-7000.
Quantity
3120
Distribution
Worldwide Distribution - USA including AL, AR, CA, CO, FL, IA, IL, KS, LA, MD, MO, MS, NC, NM, NY, OH, SC, TX & VA; and the countries of Argentina, Ecuador, Mexico, Australia, Egypt, Singapore, Belgium, Greece, Slovenia, Brazil, India, South Africa, China, Israel, United Arab Emirates, Colombia, Japan & Venezuela.
Lot, serial or model codes
Model # EC60: H44T06, J4A56P, J4A67X, H44T6E, J4A650, J4A762, J4A49A, J4A65N, J4A766, J4A49G, J4A677, J4A76F, J4A567, J4A67D, J4A76L, J4A56F, J4A67H, J4A771; Model # LONG60 lots: H44R8A , H44U3R , H44X17 , J4A55E , J4A64A; Model # SC60: H44R6C, J4A777, J4A58D, J4A77F, J4A58N, J4A78Y, J4A58P.
510(k) numbers
["K051002","K070887"]
PMA numbers
not on file
Product code
GDW
Firm FEI
1527736
Firm city
Cincinnati
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register