FDA Medical Device Recalls (openFDA)
Stryker Howmedica Osteonics Corp.: Gamma 3 System Set Screwdriver, Flexible Shaft; 4mm, This product is non-sterile Stryker Trauma GmbH, Germany; Distributed by Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430 - Orthopedic Device. The Gamma3 Locking Nail Systems are intramedullary short and long fixation nails intended for control of rotation and shortening of fractured femurs. The screwdriver is used to place the set screw in the nail.
- Recall number
- Z-1626-2008
- Recalling firm
- Stryker Howmedica Osteonics Corp.
- Product
- Gamma 3 System Set Screwdriver, Flexible Shaft; 4mm, This product is non-sterile Stryker Trauma GmbH, Germany; Distributed by Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430 - Orthopedic Device. The Gamma3 Locking Nail Systems are intramedullary short and long fixation nails intended for control of rotation and shortening of fractured femurs. The screwdriver is used to place the set screw in the nail.
- Status
- Terminated
- Initiated
- 2008-02-22
- Posted
- 2008-09-16
- Terminated
- 2010-10-25
- Reason
- Stryker Trauma has become aware that five lots of the Gamma3 Set Screwdriver have hexagon tips that were not hardened to specification. If the screwdriver tip were to deform intra-operatively, the correct positioning of the set screw would not be possible, which could lead to lateral or medial migration of the lag-screw.
- Root cause
- Finished device change control
- Action
- On February 22, 2008 Urgent Product Recall notification letters were sent to Stryker Branches/Agencies, Risk Management Departments at hospitals and surgeons via Federal Express. If you have any questions contact Rita Intorrella at (201) 831-5825 or the Hip Fracture Team at (201) 831-5291.
- Quantity
- 43 total (single and kits)
- Distribution
- Class II Recall - Nationwide Distribution.
- Lot, serial or model codes
- Product number: 1320-0231 Lot Code on Package label K409181, Lot code engraved on Screwdriver: K968226; Lot code on package label: K610297, Lot code engraved on Screwdriver: K782153; Lot code on package label; K645667, Lot code engraved on screwdriver: K533686; Lot code on package label: K688034, Lot code engraved on screwdriver: K887079; and Lot code on package label: K878962, Lot code engraved on screwdriver: K533686.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HXX
- Firm FEI
- 3003070421
- Firm city
- Mahwah
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28