FDA Medical Device Recalls (openFDA)
Hospira Inc.: LATEX-FREE; T-U-R Y-Set; Nonvented Set for Transurethral Resection Procedures; 96 inch (244 cm); For Flexible Irrigation Container Systems; List No. 6543-01; single use; Hospira Product Usage: For use in transurethral resection procedures
- Recall number
- Z-1624-2014
- Recalling firm
- Hospira Inc.
- Product
- LATEX-FREE; T-U-R Y-Set; Nonvented Set for Transurethral Resection Procedures; 96 inch (244 cm); For Flexible Irrigation Container Systems; List No. 6543-01; single use; Hospira Product Usage: For use in transurethral resection procedures
- Status
- Terminated
- Initiated
- 2014-03-25
- Posted
- 2014-05-16
- Terminated
- 2016-12-06
- Reason
- The Tyvek cover may not be completely sealed to the rest of the package allowing openings at the edge of the sterile package for the following products: List No. 04693-01, Latex Free Extension Set; List No. 06543-01, Nonvented Y-Set for Transurethral Resection; List No. 06544-01 Cystoscopy / Irrigation Set; List No. 06599-01, Large Bore Y-Irrigation Set (Latex-Free); and, List No. 15239-01, 4-Lea
- Root cause
- Process design
- Action
- Hospira sent an URGENT MEDICAL DEVICE RECALL letter dated March 25, 2014 to all affected customers. The letter included instructions for the direct accounts to: 1) check their inventories and quarantine any of the affected products; 2) complete and return the attached reply form to the fax number or e-mail address on the form; 3) inform potential users of these products of this recall; 4) return the affected products to Stericycle using the labels provided with the letter (additional labels or reply forms can be obtained by calling Stericycle at 877-377-5128, Monday - Friday, 8:00 AM - 5:00 PM EST); and, 5) if the products were further distributed, notify those customers and request that they contact Stericycle at 877-377-5128 (Monday - Friday, 8:00 AM - 5:00 PM EST). An updated URGENT MEDICAL DEVICE RECALL letter dated March 28, 2014 was sent to all affected customers. The letter contained instructions similar to those found in the March 25, 2014 letter.
- Quantity
- 1,109,321 set (US); 15,113 sets (Foreign)
- Distribution
- Worldwide Distribution - US: Nationwide in the states of (AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, UT, WA, WI, WV, WY) including District of Columbia and Puerto Rico; *** the countries of Barbados, Canada, Costa Rica, Dominican Republic, Hong Kong, Kuwait, Libya, Saudi Arabia, United Arab Emirates
- Lot, serial or model codes
- List No. 6543-04-01 (6543-01); Lot Numbers: 13106NS, 13170NS, 13251NS, 14261NS, 14262NS, 15139NS, 16117NS, 17131NS, 17132NS, 17184NS, 18110NS, 18111NS, 18182NS, 19133NS, 20122NS, 20123NS, 20124NS, 20205NS, 22146NS, 22147NS, 22148NS, 22149NS, 23103NS, 23104NS, 23105NS, 23106NS, 24156NS, 25130NS, 25131NS, 27109NS, 27253NS, 27286NS, 28136NS, 28137NS, 29094NS, 29095NS, 29096NS, 30194NS, 30195NS, 30196NS, 31042NS, 31043NS, 32229NS, 32230NS, 33138NS, 33139NS, 33140NS, 34286NS, 35117NS
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- GBX
- Firm FEI
- 3013319212
- Firm city
- Lake Forest
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28