Medtronic Spine LLC, formerly Kyphon Inc: Medtronic brand KyphX HV-R Bone Cement, a polymethyl bone cement consisting of a quantity of powder (20 g) and a volume of liquid (10 ml/9 g) mixed in a Kyphon mixer for up to two minutes. Model Number: CO1A; Manufactured by Elmdown LTD, London, England, for Medtronic Spine LLC, Sunnyvale, CA. Product is a medical device used during balloon kyphoplasty to cement structures in fractured vertebral bodies.

Recall number
Z-1624-2011
Recalling firm
Medtronic Spine LLC, formerly Kyphon Inc
Product
Medtronic brand KyphX HV-R Bone Cement, a polymethyl bone cement consisting of a quantity of powder (20 g) and a volume of liquid (10 ml/9 g) mixed in a Kyphon mixer for up to two minutes. Model Number: CO1A; Manufactured by Elmdown LTD, London, England, for Medtronic Spine LLC, Sunnyvale, CA. Product is a medical device used during balloon kyphoplasty to cement structures in fractured vertebral bodies.
Status
Terminated
Initiated
2009-11-10
Posted
2011-03-11
Terminated
2011-03-11
Reason
Product may not reach proper viscosity (dough state) and inconsistency or lack of homogeneity during cement preparation and during delivery period.
Root cause
Other
Action
The firm, Medtronic Spine LLC, sent an "URGENT DEVICE RECALL NOTICE" dated November 9, 2009 to the customers. The notice described the product, problem and actions to be taken. The customers were instructed to return the affected lots to Medtronic Spine; to not use any of the affected product that they may have received (Medtronic will replace any affected product in their inventory), the customer should contact their Medtronic Spine sales representative or customer service at [866-959-7466 or for geography] for assistance with replacement product, and complete and return the URGENT DEVICE RECALL NOTICE: Action Required form to Medtronic Spine LLC via fax at (408) 543-6192 or mail to: Medtronic Spine LLC, Attn: Recall Coordinator, 1221 Crossman Avenue, Sunnyvale, CA 94089. If you have any questions about this notification, please do not hesitate to contact Sr. Director, Regulatory Affairs via email: pamela.n.segale@medtronic.com or phone at (408) 548-5235.
Quantity
not on file
Distribution
USA including states of: AL, AR, AZ, CA, CO, DC, FL, GA, IA, IN, KY, MA, MD, MO, MT, NC, NM, NV, NY, OH, PA, SC, TN, TX, UT, VA, WI, and WV; and countries including: Argentina, Austria, Belgium, Brazil, Canada, Chile, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Ivory Coast, Jordan, Kuwait, Lebanon, Luxembourg, Lybia, Malaysia, Malta, Mexico, Morocco, Netherlands, New Caledonia, Norway, Panama, Philippines, Poland, Portugal, Puerto Rico, Reunion, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Turkey, United Arab Emirates, United Kingdom and Vietnam.
Lot, serial or model codes
Lot Numbers: EL15106, EL15406, EL15706, EL16006, EL16406, EL16506, EL16806, EL17006, EL17307, EL17507, EL17607, EL17707, EL17807, EL18007, EL18307, EL18407, EL18707, EL18907, EL19207, EL19507, EL19807, EL20007, EL20107, EL20307, EL20407, EL20507, EL20607, EL20808, EL20908, EL21008, EL21208, EL21308, EL21608, EL21708, EL21908, EL22008, EL22108, EL22308, EL22908, EL23008, EL23508, EL23608, EL23708, EL24008, 24108, EL24208, EL24408, EL24508, EL24608, EL25008, EL25108, EL25408, EL25508, EL26409, EL26509
510(k) numbers
["K033801","K041584"]
PMA numbers
not on file
Product code
NDN
Firm FEI
3004904811
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register