Skip to the content

FDA Medical Device Recalls (openFDA)

Fresenius Medical Care Holdings, Inc.: Fresenius Optiflux F250NRe Hemodialyzer, Single Use, Catalog Number: 0500325E, Advanced Fresenius Polysulfone. Intended for use as an artificial kidney system to the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration and hemodiafiltration.

Recall number
Z-1622-2011
Recalling firm
Fresenius Medical Care Holdings, Inc.
Product
Fresenius Optiflux F250NRe Hemodialyzer, Single Use, Catalog Number: 0500325E, Advanced Fresenius Polysulfone. Intended for use as an artificial kidney system to the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration and hemodiafiltration.
Status
Terminated
Initiated
2011-02-16
Posted
2011-03-22
Terminated
2013-02-12
Reason
There is an increased risk of an internal blood leak.
Root cause
Process control
Action
FMCNA notified consignees on 2/16/11 by telephone and follow-up with an Urgent FMCNA Optiflux F250NRe Recall letter via Certified Mail, return Receipt requested. The firm states in the letter that there is an increased risk of an internal blood leak affecting a small percentage of the two lots of dialyzers. They are asking customers to examine their inventory, to discontinue use, and segregate the affected product in a secure place. Customers are to contact their FMCNA Customer Service Team at 1-800-323-5188 for instructions on how to return product. Used dialyzers should not be returned. If there are any questions, Customer Service should be contacted.
Quantity
38,364 units
Distribution
Worldwide Distribution -- USA and Canada.
Lot, serial or model codes
Lot numbers:  10HU02012, Expiration Date: 06/30/2013; and  10LU02016, Expiration Date: 09/30/2013.
510(k) numbers
["K082414"]
PMA numbers
not on file
Product code
KDI
Firm FEI
3001451489
Firm city
Waltham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register