Boston Scientific Corporation: ELUVIA Drug-Eluting Vascular Stent System - Product Usage: indicated for improving luminal diameter in the treatment of symptomatic de-novo or restenotic lesions in the native superficial femoral artery (SFA) and/or proximal popliteal artery with reference vessel diameters (RVD) ranging from 4.0-6.0 mm and total lesion lengths up to 190 mm.

Recall number
Z-1621-2021
Recalling firm
Boston Scientific Corporation
Product
ELUVIA Drug-Eluting Vascular Stent System - Product Usage: indicated for improving luminal diameter in the treatment of symptomatic de-novo or restenotic lesions in the native superficial femoral artery (SFA) and/or proximal popliteal artery with reference vessel diameters (RVD) ranging from 4.0-6.0 mm and total lesion lengths up to 190 mm.
Status
Terminated
Initiated
2021-04-21
Posted
not on file
Terminated
2023-08-07
Reason
One batch of the ELUVIATM Drug Eluting Vascular Stent System incorrect placement of some 6x40mm ELUVIA stents into 6x100mm packaging.
Root cause
Reprocessing Controls
Action
On April 21, 2021 the firm sent a letter to its customers with Immediate Action Required with the following instructions: If you identify any product from the listed batch within your inventory, please segregate the product immediately and return it to Boston Scientific in accordance with the enclosed instructions. You will receive replacement product for all removed product returned to Boston Scientific. If you are a distributor, please note that the removal depth is to the hospital level and the removal notification should be forwarded to your customers.
Quantity
27 UNITS
Distribution
US Nationwide distribution in the states of AL, AR, AZ, CA, DE, HI, IL, KS, KY, MA, MD, NC, NJ, NV, NY, OH, OK, PA,PR, TX, VA,WA,WV.
Lot, serial or model codes
LOT: 25838357
510(k) numbers
not on file
PMA numbers
not on file
Product code
NIU
Firm FEI
3002095335
Firm city
Maple Grove
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register