FDA Medical Device Recalls (openFDA)
SEASPINE ORTHOPEDICS CORPORATION: Mariner RDX MIS Polyaxial Head, REF: MX1-001010, part of Mariner RDX MIS Alpha Instruments and Implants, REF: AMRDXMIS, which are pedicle screw systems.
- Recall number
- Z-1620-2024
- Recalling firm
- SEASPINE ORTHOPEDICS CORPORATION
- Product
- Mariner RDX MIS Polyaxial Head, REF: MX1-001010, part of Mariner RDX MIS Alpha Instruments and Implants, REF: AMRDXMIS, which are pedicle screw systems.
- Status
- Open, Classified
- Initiated
- 2024-02-28
- Posted
- 2024-04-23
- Terminated
- not on file
- Reason
- Screw heads, of pedicle screw systems, have partial thread within the screw head that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.
- Root cause
- Nonconforming Material/Component
- Action
- On 2/26/24, the firm phoned customers and informed them of the recall and that they would receive an email outlining details of the field action and subsequent steps to be taken. On 2/28/24, recall emails were sent to customers asking customers to respond to the emails to coordinate product return and replacement. Customers with questions can contact the firm at olivia.disanto@seaspine.com.
- Quantity
- 444 Screw Heads
- Distribution
- US: CA, FL, CO, TX
- Lot, serial or model codes
- REF/UDI-DI/Lots: MX1-001010/10889981298647/ AW161319D, AW164195D, AW166994D, AW167309D, AW167422D, AW169350F, AW169351F. AMRDXMIS/10889981303297/ AMRDXMIS-001, AMRDXMIS-002, AMRDXMIS-003, AMRDXMIS-004, AMRDXMIS-005
- 510(k) numbers
- ["K222110"]
- PMA numbers
- not on file
- Product code
- NKB
- Firm FEI
- 3012120772
- Firm city
- Carlsbad
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28